Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  • Ping various computer/network hardware to confirm they are alive

    • kaimama

    • control1

    • eds1

    • mpulizzi1

    • mpulizzi2

    • iolan1

  • Check lastalive keyword on software (Servers)

    • mosfire

    • mcsus

    • mdhs

    • mds

    • mfcs

    • mmdcs

    • mmf1s

    • mmf2s

    • mmgss

    • mmgts

    • mmprs

    • mp1s

    • mp2s

    • mpas

    • mscs

    • mtcs

  • Verify PIDs exist for software

    • write_image

    • fcstrack

    • ds9relay

    • lickserv

  • Check daemons (only if observing)

    • watchrot

    • watchslew

    • watchfcs

    • autodisplay

  • Verify power status

    • glycol_power = on

    • CSU_controller_power = on

    • CSU_drives_power = on

    • Jade2_power = on

    • computer_power = on

    • Lantronix_power = on

    • guider_focus_power = on

    • ion_pump_power = on

    • MAGIQ_guider_power = on

    • temp_control_power = on

    • motor_box_power = on

    • power_supply_power = on

    • FCS_controller_power = on

    • dewar_heater_power = on

    • CCR_1_power = on

    • CCR_2_power = on

  • Verify stages

    • dust cover, mmdcs status is OK

    • filter wheel 1, mmf1s status is OK

    • filter wheel 2, mmf2s status is OK

    • grating turret, mmgts status is OK

    • grating shim, mmgss status is OK

    • pupil rotator, mmprs status is OK or TRACKING

    • CSU, mcsus status is OK

  • Check Settings

    • glycol flow (mdhs glysup = Flow)

    • detector temp (mtcs curtmp1 = 77 +/- 0.1)

    • bulkhead temp (mtcs curtmp2 = 120 +/- 1)

    • CCR remote control (mscs chremote = 1)

    • CCR 1 Speed (mscs dacout1r = 45 +/- 1)

    • CCR 2 Speed (mscs dacout2r = 45 +/- 1)

    • window heater voltage (mdhs tmp5 = 17.5 +/- 1.5)

    • CSU electronics temp (mdhs tmp2 = 0-30)

    • cabinet temp (mdhs tmp3 = 0-30)

    • thermocouple pressure (mpas g1str = 0-1e-4)

    • ion pump 1 pressure (mpas g3str = 0 - 1e-7)

    • ion pump 2 pressure (mpas g4str = 0 - 1e-7)

    • cabinet door status (mdhs ebaydoor = Closed)

    • FCS correcting (mfcs active = 1)

    • datataking system (mds ready = 1)

    • current instrument (dcs currinst = mosfire)

...

View file
namemosfire_testAll

Instrument States

For the above checks, we need two states: one for observing and one for daytime (ready? standby? we need a good term for this). We may also want to differentiate between being on the deck and on the telescope.

  • Can we determine whether MOSFIRE is on deck or on the telescope programmatically?

  • Determine observing by combination of multiple factors: telescope has been tracking (guiding?) recently, hatch has been open recently, telescope elevation is >0, detector has been exposing recently. These are just ideas.

Mosfire Checkout (Incomplete)

...