Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Setting

When Needed

Notes

Build

Observing

Maintenance

Number of measurements 'n' to include in average.

YES

NO

MAYBE

Perhaps start with 10

Time delay 't' between measurements used for average.

YES

NO

MAYBE

Perhaps start with 0.5 seconds.

Device setup parameters follow:

Set offset subtraction

YES

NO

MAYBE

Equivalent to “ZERO” button on device.

Default KPF default should be ‘Z0’ (no offset subtraction).

Set calibration wavelength

YES

NO

MAYBE

Equivalent to wavelength buttons on device.

Default KPF default value should be 658 nm. Remote command might be 'W0658'.

Set attenuator mode

YES

NO

MAYBE

“ATTN” button on device.

Default KPF default should be set to ‘off’ (i.e. no attenuator in place). Remote command is ‘A0’

Set measurement units

YES

NO

MAYBE

Equivalent to “UNITS” button on device.

Default KPF default should be Watts (remote command ‘U1').

Set averaging mode

YES

NO

MAYBE

Equivalent to “AVG” button on device.

Default KPF default should be set to ‘Fast’ mode (no averaging by device). Remote command for ‘Fast’ mode is ‘F3’.

Set audible beep on/off

YES

NO

MAYBE

Equivalent to “BEEP” button on device.

Default KPF default should be ‘off’ (remote command ‘B0’)

Set automatic range

YES

NO

MAYBE

Equivalent to “AUTO” range on device.

Default KPF default should be auto-range (remote command ‘R0’).

Turn display on/off

YES

NO

MAYBE

Default KPF default should be ‘off’ (remote command ‘K0’). Likely device produces less heat (and isn’t a light source) with the display off.

Local lockout

YES

NO

MAYBE

I presume KPF default should be ‘locked’ while under remote operation (remote command ‘L1’)

...