Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

Table of Contents:

Table of Contents

Page Status

...

Date

...

Initials

...

Change

...

June 23, 2021

...

SG

...

Engineering modes updated, and device set up parameters (and mode settings) spreadsheet now included.

...

Introduction

...

Relevant Devices

Description

Qty

Source

Part Number

IP Address Link

Green CCD Cold Head
Temperature Controller

1

Instec

mK2000

Aux Rack IP List

Red CCD Cold Head
Temperature Controller

1

Instec

mK2000

Aux Rack IP List

Green LN2 Pump

1

Instec

LN2-SYS

N/A

Red LN2 Pump

1

Instec

LN2-SYS

N/A

...

Action

Notes

1

Read telemetry (Green)

Continuously running

2

Read telemetry (Red)

3

Operational Mode: Cold (Green)

Operational mode of CCD temperature maintained at cold temperature set point.

4

Operational Mode: Cold (Red)

5

Engineering Mode: Warm (Idle) (Green)

Engineering mode for idle case of warm CCD under no temperature control.

6

Engineering Mode: Warm (Idle) (Red)

7

Engineering Mode: Cool Down (Green)

Engineering and maintenance only. Controlled ramp of cold head temperature from warm to cold.

8

Engineering Mode: Cool Down (Red)

9

Engineering Mode: Warm Up (Green)

Engineering and maintenance only. Controlled ramp of cold head temperature from cold to warm.

10

Engineering Mode: Warm Up (Red)

11

Engineering Mode: Set PID (Green)

Engineering and maintenance only. Uploads a PID parameter table to the device.

12

Engineering Mode: Set PID (Red)

...

Operational Modes

...

Setting

When Needed

Notes

Green Channel Instec

Build

Observing

Maintenance

Green PID Parameters

YES

NO

MAYBE

Instec app can save PID parameters to a .csv file. Likely convenient to use same format. Used with this command:

Code Block
TEMP:SPID

Verified with this command:

Code Block
TEMP:GPID

Green Heating/cooling status

YES

NO

MAYBE

Values: 0=Heating Only, 1=Heating and Cooling, 2=Cooling Only

Change between different modes with

Code Block
TEMP:CHSW

command.

Green Cold Temperature Set Point

YES

NO

MAYBE

Temperature in degrees C.

Used for the following commands:

Code Block
TEMP:HOLD
Code Block
TEMP:RAMP

Green Warm Temperature Set Point

YES

NO

MAYBE

Temperature in degrees C.

Used for the following commands:

Code Block
TEMP:HOLD
TEMP:RAMP

Green Cool-down Rate

YES

NO

MAYBE

Temperature change rate of degrees C per minute.

Note this value is positive-only and the mK2000 infers the direction based on current and desired temperature.

Used for the RAMP command:

Code Block
TEMP:RAMP

Green Warm-up Rate

YES

NO

MAYBE

Temperature change rate of degrees C per minute

Note this value is positive-only and the mK2000 infers the direction based on current and desired temperature.

Used for the RAMP command:

Code Block
TEMP:RAMP

Green Power Percentage Value

YES

NO

MAYBE

Power setting in percent.

NOTE: the power percentage value on the Instec screen is in percent. However, power values passed through software are ratios. i.e., to set power to 5% the command would be:

Code Block
TEMP:RPP 0.05

and a following query

Code Block
TEMP:POW?

would return 0.050. Screen would say 5%.

Suggest using ‘units’ of percent for this variable to match display screen; and have software convert it to a ratio behind the scenes for commanding the Instec.

Red Channel Instec

Build

Observing

Maintenance

Red PID Parameters

YES

NO

MAYBE

See notes above for Green counterpart.

Red Heating/cooling status

YES

NO

MAYBE

See notes above for Green counterpart.

Red Cold Temperature Set Point

YES

NO

MAYBE

Temperature in degrees C.

See notes above for Green counterpart.

Red Warm Temperature Set Point

YES

NO

MAYBE

Temperature in degrees C.

See notes above for Green counterpart.

Red Cool-down Rate

YES

NO

MAYBE

Temperature change rate of degrees C per minute.

See notes above for Green counterpart.

Red Warm-up Rate

YES

NO

MAYBE

Temperature change rate of degrees C per minute

See notes above for Green counterpart.

Red Power Percentage Value

YES

NO

MAYBE

Power setting in percent.

See notes above for Green counterpart.

...

Parameter

Threshold

Notes

1

Cold Head Temperature Value Low (Green)

Status
colourRed
titleTBC

2

Cold Head Temperature Value High (Green)

Status
colourRed
titleTBC

3

Cold Head Temperature Value Low (Red)

Status
colourRed
titleTBC

4

Cold Head Temperature Value High (Red)

Status
colourRed
titleTBC

...

Relevant Datasheets & Manuals

...