Table of Contents:
Table of Contents |
---|
...
Introduction
The vacuum pump system consists of a turbo pump backed by a roots pump. The Leybold TurboVac pump is controlled via a Leybold Turbo.Control controller, which contains a webserver to allow control of the pump via a webpage in a browser. The Leybold EcoDry roots pump is not under remote control; within KPF we turn it on and off via a PDU outlet on the Vac Cart.
The Turbo.Control controller also allows connection of two vacuum gauges; we use both of these ports to connect a low and a high vacuum gauge. These two gauges monitor the vacuum pressure within the pipe going to the main KPF vacuum chamber. They are mounted on the Vac Cart and sit on the Vac Cart side of the main chamber gate valve.
Direct computer control of the TurboVac pump is possible via serial commands, however creating a full KTL dispatcher for the pump was deemed too difficult given the dozens of commands required, as well as the failsafes required to not damage the pump. As such we decided to use the vendor-supplied web-based software to control the pump, which has all the features we need built-in. This webpage also provides pump telemetry, and a KPF KTL dispatcher monitors this webpage and recorders telemetry values to KTL keywords.
...
Relevant Devices
Description | Qty | Source | Part Number | Notes |
---|---|---|---|---|
Turbo Pump | 1 | Leybold | LH TurboVac 450iX | |
Turbo Controller | 1 | Leybold | Turbo.Control i | |
Roots Pump | 1 | Leybold | LH EcoDry65 | Slaved to Turbo Pump (no direct control) |
High Vac gauge | 1 | Leybold | PTR 90 N | Combo Pirani and Cold Cathode gauge |
Low Vac gauge | 1 | Leybold | TTR 91 N S | Pirani gauge |
...