Relevant Functional Block Diagram

...

Software Actions

2

Action

Notes

1

Read light intensity valueSingle measurement.

That is, report the Wattage value the detector is currently reporting.

Note calibration unit translation stage has to move detector into beam first.Value

2

Calculate average of ‘n' discrete single measurements taken ’t' seconds apart.

Average several single measurements together to smooth out measurement noise/detector variability.

This value reported as telemetry (see below)

.

This approach was used for measurements at SSL using a similar device; see example Python code at bottom of this page.

...

Required Adjustable Settings

...

Setting

When Needed

Notes

Build

Observing

Maintenance

Number of measurements 'n' to include in average.

YES

NO

MAYBE

Perhaps start with 10

Time delay 't' between measurements used for average.

YES

NO

MAYBE

Perhaps start with 0.5 seconds.

Device setup parameters follow:

Set offset subtraction

YES

NO

MAYBE

Equivalent to “ZERO” button on device.

Default should be ‘Z0’ (no offset subtraction).

Set calibration wavelength

YES

NO

MAYBE

Equivalent to wavelength buttons on device.

Default value should be 658 nm. Remote command might be 'W0658'.

Set attenuator mode

YES

NO

MAYBE

“ATTN” button on device.

Default should be set to ‘off’ (i.e. no attenuator in place). Remote command is ‘A0’

Set measurement units

YES

NO

MAYBE

Equivalent to “UNITS” button on device.

Default should be Watts (remote command ‘U1').

Set averaging mode

YES

NO

MAYBE

Equivalent to “AVG” button on device.

Default should be set to ‘Fast’ mode (no averaging by device). Remote command for ‘Fast’ mode is ‘F3’.

Set audible beep on/off

YES

NO

MAYBE

Equivalent to “BEEP” button on device.

Default should be ‘off’ (remote command ‘B0’)

Set automatic range

YES

NO

MAYBE

Equivalent to “AUTO” range on device.

Default should be auto-range (remote command R0‘R0’).

Turn display on/off

YES

NO

MAYBE

Default should be ‘off’ (remote command ‘K0’). Likely device produces less heat (and isn’t a light source) with the display off.

Local lockoutDefault

YES

NO

MAYBE

I presume default should be ‘locked’ while under remote operation (remote command ‘L1’)

...

Parameter

Threshold

Notes

1

Calibration source Source 1 intensity low

TBD

2

Calibration Source 2 intensity low

TBD

3

Calibration Source 3 intensity low

TBD

4

Calibration Source 4 intensity low

TBD

5

Calibration Source 5 intensity low

TBD

6

Calibration Source 6 intensity low

TBD

7

Calibration Source 7 intensity low

TBD

8

Calibration Source 8 intensity low

TBD

...

Relevant Datasheets & Manuals

...