Introduction
Each cryostat will have two ion pumps attached. One will be operated continuously and the second will sit idle as a backup. Each pump has a dedicated controller (for a total of four in the system).
...
Proposal: name each ion pump with respect to its camera (Green/Red) and the side of the spectrometer it is on (Echelle/Collimator)
...
Relevant Devices
Description | Qty | Source | Part Number |
---|---|---|---|
Ion Pump Controller | 4 | Gamma | SPC-1-P-S-1-US110-E-S-N |
Ion Pump | 4 | Gamma | 3SDI1V5KNN (3S TiTan) |
Air Cables | 4 | Gamma | 10K-SC10-10K (10 m) |
Vacuum Cables | 4 | Custom |
Note there are two pumps per cryostat. One will be operated and the second will sit idle as a backup.
Each pump has a dedicated controller.
- |
...
Relevant Functional Block Diagram
...
Software Actions
Action | Notes | |||||
---|---|---|---|---|---|---|
1 | SG needs to update | 2 | 3 | Green controller 1: take pressure measurement | Value GRN_ION_E: Read telemetry | Pressure, Voltage, and Current values reported as telemetry (see below) |
4 | Green controller 2: take pressure measurement | Value reported as telemetry (see below) | ||||
5 | Red controller 1: take pressure measurement | Value reported as telemetry (see below) | ||||
6 | Red controller 2: take pressure measurement | Value reported as telemetry (see below) | ||||
7 | Green controller 1: turn pump on | 8 | Green controller 12 | GRN_ION_C: Read telemetry | ||
3 | RED_ION_E: Read telemetry | |||||
4 | RED_ION_C: Read telemetry | |||||
5 | GRN_ION_E: turn pump on | These controls should have a lock applied so they cannot be accidentally switched. | ||||
6 | GRN_ION_E: turn pump off | |||||
9 | Green controller 27 | GRN_ION_C: turn pump on | 10 | Green controller 2|||
8 | GRN_ION_C: turn pump off | |||||
11 | Red controller 19 | RED_ION_E: turn pump on | 12 | Red controller 1|||
10 | RED_ION_E: turn pump off | |||||
13 | Red controller 211 | RED_ION_C: turn pump on | 14 | Red controller 2|||
12 | RED_ION_C: turn pump off |
...
Required Adjustable Settings
...
Parameter | Report Frequency | Purpose | Notes | |
---|---|---|---|---|
1 | SG needs to update | |||
2 | ||||
3 | Green controller 1 pressure value | Once per minute (TBC) | Monitor green cyrostat pressure | |
4 | Green controller 2 pressure value | Once per minute (TBC) | Monitor green cyrostat pressure | |
5 | Red controller 1 pressure value | Once per minute (TBC) | Monitor red cyrostat pressure | |
6 | Red controller 2 pressure value | Once per minute (TBC) | Monitor red cyrostat pressure |
...
Parameter | Threshold | Notes | |
---|---|---|---|
1 | Green controller 1 pressure high | TBC | Only one pump per cyrostat will be in operation at any time. The alarm system would need to know which pump/controller is in operation at any time to monitor the correct pressure value. |
2 | Green controller 2 pressure high | TBC | |
3 | Red controller 1 pressure high | TBC | |
4 | Red controller 2 pressure high | TBC | |
5 | |||
6 | High Voltage turns off | ||
7 | |||
8 | |||
9 |
...
Relevant Datasheets & Manuals
...
Device Command List for KPF
Noted as a ‘draft’ at this point as need to finalize Arc Detection command setting.
View file | ||
---|---|---|
|
...