Page Status

DRAFT/INCOMPLETE

...

Etalon Status Update

Based on C. Schwarb’s email on Nov 4, 2022.

  •  Need update from Julian Sturmer regarding temperature controllers and etalon computer setup.
  •  Set up etalon computer as a vendor-supplied machine, similar to the Menlo machine (question)

Dwight and John V. may perform the first two tasks described below in the week of Nov 14, 2022, if they have time.

  •  Backfill Dewar – backfill the etalon with Helium, at about atmospheric pressure

The Etalon is currently pumped and closed. The valve I used is the final version, and all metal seal valves can only be closed a small number of times. It has indicator marks, and every time you close it, you turn it another half indicator mark tighter than the last time. Currently, it sits on the very first mark. The manual of that valve is attached. Check pages 6 and 7. 

On the exit port of the valve, I installed a CF16 to KF40 adapter, to make it interfaceable with the rest of your stuff. There is a KF40 Tee somewhere in the basement, and I know you have more of them, and also KF40 O-ring sealed valves. I suggest adding a Tee like last time to interface to a pump and the helium bottle, and possibly one of those O-ring KF valves in front (towards etalon) of it, so you can easily close the chamber off once it's filled with He and "burped". The all-metal valve needs a bit of care to close, can't do it quickly.

View file
nameBP4A01E-20190918-Right_angle_all_metal_bakeable_valve_instruction_manual.pdf

  •  Mount Sidewall – close the sidewall, and put it back onto the bench in an orientation that allows the cables to reach the temperature controllers

Self-explanatory, and the screws (plastic) are stuck to the panel with sticky tape.  If you can, would you put some sticky tape over the gaps in the corners all around?

  •  Install Temperature Controllers – plug those cables into the vescent temperature controllers (order below)

Because Julian did not get access to the control computer until very recently, these are not tuned up, which we need to really get on. The controllers are the two Vescent units sitting side by side in the rack on the right side of the cal bench. Cables go into the back - suggest pulling one controller out and having a look at how the channels are numbered. 

The cables from the etalon are all labeled, as below. 

Vescent 1 (left one):

CH1: Etalon 10k        (10k thermistor glued into the etalon mount)

CH2: Bottom Int         (bottom of the active shield)

CH3: Shield               (tube of the active shield)

CH4: Top Int              (top of the active shield)

Vescent 2:

CH1: Passive 10k      (bottom of the passive shield)

CH2: External            (external chamber heaters, all connected)

...

Etalon Rack Devices

Description

Qty

Manufacturer

Part Number

Voltage
[V]

Current
[A]

Power
[VA]

PDU Plugs

Ethernet
Req'd?

Rack U’s Req’d
(1U = 1.75”)

Controlling Software
[Etalon/KPF]

WMKO Heat Exchanger

1

WMKO

-

1

-

6

N/A

Etalon Rack Network Switch

1

Aruba

2530-24G

1

-

1

N/A

Etalon Rack PDU

1

Eaton

EMAT08-10

-

1

1

Etalon computer through web interface

Etalon White Light Source

1

NKT Photonics

SuperK EVO HP EU-4

4

Etalon?

Etalon digital acquisition interface

2

MCDAQ

USB-1808X

USB

0

-

1 (shared)

Etalon

Laser current controller

1

Custom

-

110

60

1

-

3

-

Temperature controller

2

Vescent

SLICE-QT

110

50

1

-

3 (for both)

Etalon

Ion getter pump & controller

1

Gamma

Gamma SPC-1-P-S-1-US110-E-S-N

110

0.04

1

1

3

(shared?)

Etalon

Vacuum gauge

1

Thyracont

VSM79

110

5

1

-

-

Etalon

Etalon computer

1

InoNet (TBD)

MAYFLOWER-ID/i7-8700Rp

110

250

2

1

2

Etalon

Notes:

  • For “Controlling Software” column:

    • “Etalon” = device is only controlled by software on etalon PC;

    • “KPF” = device needs to be controlled by KPF instrument control software

  • For “Etalon computer”:

    • any 19” server with Linux OS (preferred: Ubuntu) with at least 4 cores. >=1TB SSD, >=2TB HDD, >=16GB RAM will work

    • redundant power supply recommended (included in table)

    • necessary interfaces (TBD): 4x USB 2.0, 2x USB 3.0

...

Other Facility Services

Service

Description

Ethernet

Ethernet line from the KPF main switch to the etalon rack switch.

Compressed Air

To float the optical table shared with the calibration unit.

Glycol

For the WMKO heat exchanger in the top of the etalon rack.

Dry Air

?

Other?

?

...

Relevant Functional Block Diagram

...

List of etalon actions that the KPF instrument control software will need to perform (i.e. move something, read a value, measure something, etc.)

Action

Notes

1

2

3

Email exchange between Julian and Will about KTL access to etalon telemetry:

...

List of etalon device settings that the KPF instrument control software would need to control or change.

Setting

When Needed
[Yes/No/Maybe]

Notes

-

Build

Observing

Maintenance

-

...

Required Telemetry

List of etalon telemetry that the KPF instrument control software should monitor, save, or make available for image headers.

Parameter

Report Frequency

Purpose

Notes

1

2

...

Required Alarms

Etalon telemetry items that should have a KPF instrument software alarm raised when telemetry passes a threshold value.

Parameter

Threshold

Notes

1

2

...

Relevant Datasheets & Manuals

Description

File

Source

1

2