Proposal: name each ion pump with respect to its camera (Green/Red) and the side of the spectrometer it is on (Echelle/Collimator). The driver here is that the name then gives the location of the pump (whereas labels like “Green #1” or “Red 10 o’clock” or “Green Prime” require more information to derive the position).
However: Brad - please let me know if this is too cumbersome!
...
Relevant Devices
Description | Qty | Source | Part Number |
---|---|---|---|
Ion Pump Controller | 4 | Gamma | SPC-1-P-S-1-US110-E-S-N |
Ion Pump | 4 | Gamma | 3SDI1V5KNN (3S TiTan) |
Air Cables | 4 | Gamma | 10K-SC10-10K (10 m) |
Vacuum Cables | 4 | Custom | - |
...
Software Actions
Action | Notes | |||||||
---|---|---|---|---|---|---|---|---|
1 | GRN_ION_E: Read telemetry | Pressure, Voltage, Current and Current High Voltage Status values reported as telemetry (see below) | ||||||
2 | GRN_ION_C: Read telemetry | |||||||
3 | RED_ION_E: Read telemetry | |||||||
4 | RED_ION_C: Read telemetry | |||||||
5 | GRN_ION_E: turn pump on
| These controls should have a lock applied so they cannot be accidentally switched. | ||||||
6 | GRN_ION_E: turn pump off
| |||||||
7 | GRN_ION_C: turn pump on
| |||||||
8 | GRN_ION_C: turn pump off
| |||||||
9 | RED_ION_E: turn pump on
| |||||||
10 | RED_ION_E: turn pump off
| |||||||
11 | RED_ION_C: turn pump on
| |||||||
12 | RED_ION_C: turn pump off
|
...
Required Adjustable Settings
Setting | When Needed | Notes | ||
---|---|---|---|---|
Build | Observing | Maintenance | ||
...
Required Telemetry
Pump | Parameter | Report Frequency | Purpose | Notes | ||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | SG needs to update | 2 | 3 | Green controller 1 pressure value | Once per minute (TBC) | Monitor green cyrostat pressure | 4 | Green controller 2 pressure value | Once per minute (TBC) | Monitor green cyrostat pressure | 5 | Red controller 1 pressure value | Once per minute (TBC) | Monitor red cyrostat pressure | 6 | Red controller 2 pressure value | Once per minute (TBC) | Monitor red cyrostat pressure | GRN_ION_E | Pressure | Once per minute
| Monitor cyrostat pressure | ||||||||
2 | Voltage | Once per minute
| Monitor cyrostat pressure | |||||||||||||||||||||||||||
3 | Current | Once per minute
| Monitor cyrostat pressure | |||||||||||||||||||||||||||
4 | High Voltage Status | Once per minute
| Ensure pump is on | |||||||||||||||||||||||||||
5 | GRN_ION_C | Pressure | Once per minute
| Monitor cyrostat pressure | ||||||||||||||||||||||||||
6 | Voltage | Once per minute
| Monitor cyrostat pressure | |||||||||||||||||||||||||||
7 | Current | Once per minute
| Monitor cyrostat pressure | |||||||||||||||||||||||||||
8 | High Voltage Status | Once per minute
| Ensure pump is on | |||||||||||||||||||||||||||
9 | RED_ION_E | Pressure | Once per minute
| Monitor cyrostat pressure | ||||||||||||||||||||||||||
10 | Voltage | Once per minute
| Monitor cyrostat pressure | |||||||||||||||||||||||||||
11 | Current | Once per minute
| Monitor cyrostat pressure | |||||||||||||||||||||||||||
12 | High Voltage Status | Once per minute
| Ensure pump is on | |||||||||||||||||||||||||||
13 | GRN_ION_C | Pressure | Once per minute
| Monitor cyrostat pressure | ||||||||||||||||||||||||||
14 | Voltage | Once per minute
| Monitor cyrostat pressure | |||||||||||||||||||||||||||
15 | Current | Once per minute
| Monitor cyrostat pressure | |||||||||||||||||||||||||||
16 | High Voltage Status | Once per minute
| Ensure pump is on |
...
Required Alarms
Parameter | Threshold | Notes | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | Green controller 1 pressure high | TBCGRN_ION_E pressure too high |
| Only one pump per cyrostat will be in operation at any time. The alarm system would need to know which pump/controller is in operation at any time to monitor the correct pressure value. | ||||||||
2 | Green controller 2 pressure high | TBC | ||||||||||
3 | Red controller 1 pressure high | TBC | ||||||||||
4 | Red controller 2 pressure high | TBC | ||||||||||
5 | 6 | High Voltage turns off | 7 | 8 | 9GRN_ION_C pressure too high |
| ||||||
3 | RED_ION_E pressure too high |
| ||||||||||
4 | RED_ION_C pressure too high |
| ||||||||||
5 | GRN_ION_E high voltage off |
| ||||||||||
6 | GRN_ION_C high voltage off |
| ||||||||||
7 | RED_ION_E high voltage off |
| ||||||||||
8 | RED_ION_C high voltage off |
|
...
Relevant Datasheets & Manuals
...
Device Command List for KPF
Noted as a ‘draft’ at this point as I need to finalize Arc Detection command setting.