Comment: First solid draft; ready for Brad to get started with.

...

Proposal: name each ion pump with respect to its camera (Green/Red) and the side of the spectrometer it is on (Echelle/Collimator). The driver here is that the name then gives the location of the pump (whereas labels like “Green #1” or “Red 10 o’clock” or “Green Prime” require more information to derive the position).
However: Brad - please let me know if this is too cumbersome!

...

Relevant Devices

Description

Qty

Source

Part Number

Ion Pump Controller

4

Gamma

SPC-1-P-S-1-US110-E-S-N

Ion Pump

4

Gamma

3SDI1V5KNN (3S TiTan)

Air Cables

4

Gamma

10K-SC10-10K (10 m)

Vacuum Cables

4

Custom

-

...

Software Actions

Action

Notes

1

GRN_ION_E: Read telemetry

Pressure, Voltage, Current and Current High Voltage Status values reported as telemetry (see below)

2

GRN_ION_C: Read telemetry

3

RED_ION_E: Read telemetry

4

RED_ION_C: Read telemetry

5

GRN_ION_E: turn pump on

Status
colourGreen
titleON

These controls should have a lock applied so they cannot be accidentally switched.

6

GRN_ION_E: turn pump off

Status
colourRed
titleOFF

7

GRN_ION_C: turn pump on

Status
colourGreen
titleON

8

GRN_ION_C: turn pump off

Status
colourRed
titleOFF

9

RED_ION_E: turn pump on

Status
colourGreen
titleON

10

RED_ION_E: turn pump off

Status
colourRed
titleOFF

11

RED_ION_C: turn pump on

Status
colourGreen
titleON

12

RED_ION_C: turn pump off

Status
colourRed
titleOFF

...

Required Adjustable Settings

Setting

When Needed

Notes

Build

Observing

Maintenance

...

Required Telemetry

Pump

Parameter

Report Frequency

Purpose

Notes

1

SG needs to update

23

Green controller 1 pressure value

Once per minute (TBC)

Monitor green cyrostat pressure

4

Green controller 2 pressure value

Once per minute (TBC)

Monitor green cyrostat pressure

5

Red controller 1 pressure value

Once per minute (TBC)

Monitor red cyrostat pressure

6

Red controller 2 pressure value

Once per minute (TBC)

Monitor red cyrostat pressure

GRN_ION_E

Pressure

Once per minute

Status
colourRed
titleTBC

Monitor cyrostat pressure

2

Voltage

Once per minute

Status
colourRed
titleTBC

Monitor cyrostat pressure

3

Current

Once per minute

Status
colourRed
titleTBC

Monitor cyrostat pressure

4

High Voltage Status

Once per minute

Status
colourRed
titleTBC

Ensure pump is on

5

GRN_ION_C

Pressure

Once per minute

Status
colourRed
titleTBC

Monitor cyrostat pressure

6

Voltage

Once per minute

Status
colourRed
titleTBC

Monitor cyrostat pressure

7

Current

Once per minute

Status
colourRed
titleTBC

Monitor cyrostat pressure

8

High Voltage Status

Once per minute

Status
colourRed
titleTBC

Ensure pump is on

9

RED_ION_E

Pressure

Once per minute

Status
colourRed
titleTBC

Monitor cyrostat pressure

10

Voltage

Once per minute

Status
colourRed
titleTBC

Monitor cyrostat pressure

11

Current

Once per minute

Status
colourRed
titleTBC

Monitor cyrostat pressure

12

High Voltage Status

Once per minute

Status
colourRed
titleTBC

Ensure pump is on

13

GRN_ION_C

Pressure

Once per minute

Status
colourRed
titleTBC

Monitor cyrostat pressure

14

Voltage

Once per minute

Status
colourRed
titleTBC

Monitor cyrostat pressure

15

Current

Once per minute

Status
colourRed
titleTBC

Monitor cyrostat pressure

16

High Voltage Status

Once per minute

Status
colourRed
titleTBC

Ensure pump is on

...

Required Alarms

9

Parameter

Threshold

Notes

1

Green controller 1 pressure high

TBCGRN_ION_E pressure too high

Status
colourRed
titleTBD

Only one pump per cyrostat will be in operation at any time.

The alarm system would need to know which pump/controller is in operation at any time to monitor the correct pressure value.

2

Green controller 2 pressure high

TBC

3

Red controller 1 pressure high

TBC

4

Red controller 2 pressure high

TBC

56

High Voltage turns off

78

GRN_ION_C pressure too high

Status
colourRed
titleTBD

3

RED_ION_E pressure too high

Status
colourRed
titleTBD

4

RED_ION_C pressure too high

Status
colourRed
titleTBD

5

GRN_ION_E high voltage off

Status
colourRed
titleTBD

6

GRN_ION_C high voltage off

Status
colourRed
titleTBD

7

RED_ION_E high voltage off

Status
colourRed
titleTBD

8

RED_ION_C high voltage off

Status
colourRed
titleTBD

...

Relevant Datasheets & Manuals

...

Device Command List for KPF

Noted as a ‘draft’ at this point as I need to finalize Arc Detection command setting.

View file
nameIon pump controller commands - DRAFT.xlsx

...