Ca HK Detector Stuck
Symptom
The HK detector is stuck in the exposing state.
or
The kpfexpose EXPOSE status is stuck and kpfexpose.EXPLAINNR
is "hk:ACQPHASE=wait".
Problem
Some combination of hardware and software is stuck.
Solution
Abort the existing HK exposure:
modify -s kpf_hk expose=abort
then restart the kpf_hk service:
kpf restart kpf_hk
Take test exposures to see if the system has recovered.
If the system is still sticking after test exposures, power cycle the Andor Camera, and the HK Galil on power strip J, ports 1, 2, and 5:
...
Then restart kpf_hk again:
kpf restart kpf_hk