Setting SCRIPTSTOP to “Yes” will request that the running script terminate in an orderly fashion. This may take several minutes depending on when the running script reaches a sensible breakpoint. It is important to use SCRIPTSTOP to halt a script (instead of hitting ctrl-c
) because cleanup actions will be performed after a SCRIPTSTOP (e.g. turning off laps to preserve their lifetime).
...
The agitator can be reenabled by simply setting the keyword to “Yes”. This should only be done by WMKO staff and should only be done if the agitator is fully functional. A broken or misbehaving agitator mechanism presents a significant danger to the science fibers.
SlewCal or Simultaneous Calibration Source is Wrong
Symptom
With the simultaneous calibration source (which is printed to the start of night log) or the slew cal source (visible in the OB GUI) are not the desired value.
Problem
These two sources are set from the same place and are always kept to the same value using the kpfconfig.SIMULCALSOURCE
keyword.
Solution
This value should only be changed by WMKO staff. The choice of calibration source is influenced by our desire to maintain the lifetime of the various lamps and calibration sources and by the need for warmup times on certain sources. This is not something the user should adjust, contact a WMKO Staff Astronomer if you feel it is set incorrectly.
Guider
Target is not Visible in Guider
...