The two CCDs within the main spectrometer are maintained at -100C to minimize thermal noise. Liquid nitrogen (LN2) is used to cool the CCDs to this temperature. Within KPF the temperature stability of the CCDs is of great importance, as any movements of the detector from thermal transients will appear as a movement of the spectrum. Within the science data this movement will mimic the Doppler shifts we are aiming that KPF is designed to measure.

As shown in the overview diagram just below, CCD thermal control is achieved with a multi-stage system, starting with the coarsest control being at the Instec cold head (Stage 1) and progressing though finer and finer control to the CCD (Stage 4). Stage 1 provides cooling and heating control, whereas Stages 2 -to 4 are heating control only. Each stage has a PID loop control for fine-tuning its characteristics.

In addition, there is a heater on the cyrostat body. Although this is not directly in-line with the CCD thermal path (Instec cold head Cold Head → Cold Finger → CCD), it is important as it maintains the cryostat body at a consistent temperature and hence controls the radiative environment surrounding the CCD. The cryostat body heater also has a PID loop for controlling its characteristics.

See below for details on the sensors and heaters (including KTL keyword names for each).

...

Sensor Details (and KTL Keywords)

...

Connector Number

Lakeshore Channel

KTL Keyword

1

Green Lakeshore D2

CRYOSTAT_G1

Within cryostat; likely on detector mounting plate to one side of CCD (opposite side to CRYOSTAT_G3)

2

Green Lakeshore D3 (not functioning)

CRYOSTAT_G2

Within cryostat; likely at the detector end of the cold strap

3

Green Lakeshore D4

CRYOSTAT_G3

Within cryostat; likely on detector mounting plate to one side of CCD (opposite side to CRYOSTAT_G1)

...

Connector Number

Lakeshore Channel

KTL Keyword

1

Red Lakeshore D2

CRYOSTAT_R1

Within cryostat; likely on detector mounting plate to one side of CCD (opposite side to CRYOSTAT_R3)

2

Red Lakeshore D3 (not functioning)

CRYOSTAT_R2

Within cryostat; likely at the detector end of the cold strap

3

Red Lakeshore D4

CRYOSTAT_R3

Within cryostat; likely on detector mounting plate to one side of CCD (opposite side to CRYOSTAT_R1)

...

Heater Details (and KTL Keywords)

...

Cable Details

Description

Details

Notes

Instec Sensor and Heater Cable

View file
nameDB15-Connector-Pinout Diagram.pdf

Custom 30-ft long DB15 cable made by Instec.

This goes between the Instec Controller and the vacuum feedthru on the main vacuum chamber. Within the chamber there is a second cable between the vacuum feedthru and the Instec Cold Head, however that cable is hard-wired into the Cold Head and not removable.

Cryostat Harness

View file
nameCryostat_Harnesses Rev 02.pptx

Description of the various sensor and heater cables that go to the cryostat.