Introduction

A gate valve with a 6-inch port is installed between the vacuum pumps on the Vac Cart , and the main vacuum chamber. The model chosen is pneumatically operated; it remains open as long as it is connected to compressed airactuated and can be opened or closed via an electrically-operated solenoid valve. There are also limit switches on the actuator which provide feedback on the open or closed state of the valve.

Note that if power is disconnected, or if the compressed air pressure drops, the gate will automatically close. This is desirable to help maintain the vacuum pressure within the main vacuum chamber during a power failure.

...

Relevant Devices

Description

Qty

Source

Part Number

Gate Valve

1

VAT

VAT Valve 09144−PE44

...