Lamp Type | Vendor | Serial number | Notes |
---|---|---|---|
(add etalon’s super-K information) | |||
...
Hollow Cathode Lamps
NOTE: When a lamp is changed, please record the change in the table below, and also update the appropriate KTL keyword with the new serial number. The relevant keywords for the hollow cathode lamps are: | |||
kpflamp.TH_DAILY_SN | kpflamp.TH_GOLD_SN | kpflamp.U_DAILY_SN | kpflamp.U_GOLD_SN |
...
Date | Change |
---|---|
SSL | Photron HJE0762 (which was actually Th-Ne) |
Aug 26, 2022 | Installed HJE0762 as ThAr Daily |
Dec 12, 2022 | Removed HNC0486 as ThAr Gold |
Dec 15, 2022 | Removed Green Scientific L82979 from ThAr Daily (broken) |
Feb 11, 2023 | ThAr Daily (please verify serial number) reported as having substantially reduced flux, probably starting around Jan 26 2023. |
Feb 15, 2023 | Brett installed Green Scientific L82884 in Th-Ar Daily. No flux seen in cals during subsequent days. |
Feb 21, 2023 |
(Note: the |
April 10, 2023 | Grant Hill made the change:
(Note: the |
August 24, 2023 | John Valliant installed L82932 (s/n keyword also updated to this value) |
Thorium-Argon Gold
Green Scientific part number: G858A
...