Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Here is a list of the actions that make up the MOSFIRE testAll

  • Status
    colourRed
    titleNOT a KEYWORD
    Ping various computer/network hardware to confirm they are alive

    • kaimama

    • control1

    • eds1

    • mpulizzi1

    • mpulizzi2

    • iolan1, control1, eds1, mpulizzi1, mpulizzi2, iolan1

  • Status
    colourGreen
    titleIS A KEYWORD
    Check lastalive keyword on software (Servers)

    • mosfire, mcsus, mdhs, mds, mfcs, mmdcs, mmf1s, mmf2s, mmgss, mmgts

    • mmprs

    • mp1s

    • mp2s

    • mpas

    • mscs

    • mtcs, mmprs, mp1s, mp2s, mpas, mscs, mtcs

  • Status
    colourRed
    titleNOT A KEYWORD
    Verify PIDs exist for software

    • write_image

    • fcstrack

    • ds9relay

    • lickserv, fcstrack, ds9relay, lickserv

  • Status
    colourGreen
    titleIS A KEYWORD
    Check daemons (only if observing)

    • watchrot

    • watchslew

    • watchfcs

    • autodisplay, watchslew, watchfcs, autodisplay

  • Status
    colourGreen
    titleIS A KEYWORD
    Verify power status

    • glycol_power = on, CSU_controller_power = on, CSU_drives_power = on, Jade2_power = on, computer_power = on, Lantronix_power = on, guider_focus_power = on, ion_pump_power = on, MAGIQ_guider_power = on, temp_control_power = on, motor_box_power = on, power_supply_power = on, FCS_controller_power = on, dewar_heater_power = on, CCR_1_power = on, CCR_2_power = on

  • Status
    colourGreen
    titleIS A KEYWORD
    Verify stages

    • dust cover, : mmdcs status is OK, filter wheel 1, : mmf1s status is OK, filter wheel 2, : mmf2s status is OK, grating turret, : mmgts status is OK, grating shim, : mmgss status is OK, pupil rotator, : mmprs status is OK or TRACKING, CSU, : mcsus status is OK

  • Status
    colourGreen
    titleIS a KEYWORD
    Check Settings

    • glycol flow (mdhs glysup = Flow)

    • detector temp (mtcs curtmp1 = 77 +/- 0.1)

    • bulkhead temp (mtcs curtmp2 = 120 +/- 1)

    • CCR remote control (mscs chremote = 1)

    • CCR 1 Speed (mscs dacout1r = 45 +/- 1)

    • CCR 2 Speed (mscs dacout2r = 45 +/- 1)

    • window heater voltage (mdhs tmp5 = 17.5 +/- 1.5)

    • CSU electronics temp (mdhs tmp2 = 0-30)

    • cabinet temp (mdhs tmp3 = 0-30)

    • thermocouple pressure (mpas g1str = 0-1e-4)

    • ion pump 1 pressure (mpas g3str = 0 - 1e-7)

    • ion pump 2 pressure (mpas g4str = 0 - 1e-7)

    • cabinet door status (mdhs ebaydoor = Closed)

    • FCS correcting (mfcs active = 1)

    • datataking system (mds ready = 1)

    • current instrument (dcs currinst = mosfire)

...