Verification (formerly testAll)
Here is a list of the actions that make up the MOSFIRE testAll
...
View file | ||
---|---|---|
|
Mosfire Checkout
Confirm the physical drive angle. It should not be within 10 degrees of a
multiple of 180 degreesStart the observing software as moseng or the account for the night
Check that the dark filter is selected. If not select it
Check mechanism status: If any of the mechanisms have a big red X on it,
you will need to home mechanisms. Note, if filter wheel is at the home
position, Status will be "OK," position will be "HOME", target will be
"unknown", and there will still be a big red X.Acquire an exposure
Inspect the dark image
If normal checkout:
Open mask
Image and confirm
Initialize CSU: modify -s mosfire csuinitbar=0
Image and confirm
Form 0.7x46 long slit
Image and confirm
If quick checkout
Form an 2.7x46 long slit
Image and confirm
Form an 0.7x46 long slit
Image and confirm
With the hatch closed change the observing mode to J-imaging, verify
mechanisms are okQuick Dark
Message user to verify sidecar logging