Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Table of Contents:

Table of Contents
minLevel1
maxLevel2
include
outlinefalse
indent
excludeTable of Contents:
typelist
class
printablefalse

...

In addition, there is a heater on the cyrostat body. Although this is not directly in-line with the CCD thermal path (Instec Cold Head → Cold Finger → CCD), it is important as it maintains the cryostat body at a consistent temperature and hence controls the radiative environment surrounding the CCD. The cryostat body heater also has a PID loop for controlling its characteristics.

...

...

See CCD Cold Head Temperature Controllers for more details on the electrical connections for the Instec system.

See Archon CCD Controllers for more details on the electrical connections for the Archons.

...

Sensor Details (and KTL Keywords)

Cryostat Thermal Sensor Locations

Sensor locations and KTL keywords are shown below.

...

See Thermal Sensor Locations (Spectrometer and Hallway) for the entire list of Lakeshore sensors .

The connectors on the various Lakeshore cable harnesses are labelled 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12.

This diagram indicates how they connect to the Lakeshore rear panel:

...

Green Lakeshore (cryostat sensors only)

...

Connector Number

...

Lakeshore Channel

...

KTL Keyword

...

1

...

Green Lakeshore D2

...

CRYOSTAT_G1

Within cryostat; likely on detector mounting plate to one side of CCD (opposite side to CRYOSTAT_G3)

...

2

...

Green Lakeshore D3 (not functioning)

...

CRYOSTAT_G2

Within cryostat; likely at the detector end of the cold strap

...

3

...

Green Lakeshore D4

...

CRYOSTAT_G3

Within cryostat; likely on detector mounting plate to one side of CCD (opposite side to CRYOSTAT_G1)

Red Lakeshore (cryostat sensors only)

...

Connector Number

...

Lakeshore Channel

...

KTL Keyword

...

1

...

Red Lakeshore D2

...

CRYOSTAT_R1

Within cryostat; likely on detector mounting plate to one side of CCD (opposite side to CRYOSTAT_R3)

...

2

...

Red Lakeshore D3 (not functioning)

...

CRYOSTAT_R2

Within cryostat; likely at the detector end of the cold strap

...

3

...

Red Lakeshore D4

...

CRYOSTAT_R3

Within cryostat; likely on detector mounting plate to one side of CCD (opposite side to CRYOSTAT_R1)

and details of how these sensors connect to the Green and Red Lakeshores.

...

Heater Details (and KTL Keywords)

...

Cable Details

Description

Details

Notes

Instec Sensor and Heater Cable

View file
nameDB15-Connector-Pinout Diagram.pdf

Custom 30-ft long DB15 cable made by Instec.

This goes between the Instec Controller and the vacuum feedthru on the main vacuum chamber. Within the chamber there is a second cable between the vacuum feedthru and the Instec Cold Head, however that cable is hard-wired into the Cold Head and not removable.

Cryostat HarnessCryostat Interconnect Drawing

Image Added

Green Interconnect Diagram shown, Red Cryostat is equivalent with “G” replaced by “R” in cable IDs.

Cryostat Harness Details

View file
nameCryostat_Harnesses Rev 02.pptx

Description of the various sensor and heater cables that go to the cryostat.

...