Introduction

Three Lakeshore controllers are used for high precision temperature telemetry within KPF.

...

Relevant Devices

Description

Qty

Source

Part Number

High Precision Temperature Sensors

Green Lakeshore

1

Lakeshore

224-12

Red Lakeshore

1

Lakeshore

224-12

Cryostat Temperature Sensors

Spectrometer Lakeshore

1

Lakeshore

224-12

RTD Sensors

36

Honeywell

Honeywell PN: HRTS-5760-B-U-1-12
DigiKey PN: 480-6183-ND

NOTE: we have two three identical devices. The only difference in the control software between the two will be the look-up table which maps the Lakeshore sensor input names (i.e. ‘C1’) to physical sensor locations within KPF (i.e. ‘Echelle’).same KTL dispatcher is used for all three, however each device has a its own configuration file as described within the “Dispatcher Configuration File” section below.

...

Relevant Functional Block Diagram

...

Both The three Lakeshores will be set up the same, with sensors placed at within (or withinaround) the vacuum chamber.

Software Actions

...

Action

...

Notes

...

Read 12 temperature values

...

Values reported as telemetry (see below)

...

Required Adjustable Settings

...

Setting

...

When Needed

...

Notes

...

Build

...

Observing

...

Maintenance

...

Temperature measurement frequency

...

Yes

...

No

...

Maybe

Device Set-up Parameters

Annotated command list with KPF settings.

View file
nameLakeshore 224 Command List.xlsx

Required Telemetry

...

Parameter

...

Report Frequency

...

Purpose

...

Notes

...

Input A Temperature [deg C]

...

Once every 5 sec [TBC]

...

Environmental monitoring

...

Input B Temperature [deg C]

...

Input C1 Temperature [deg C]

...

Input C2 Temperature [deg C]

...

Input C3 Temperature [deg C]

...

Input C4 Temperature [deg C]

...

Input C5 Temperature [deg C]

...

Input D1 Temperature [deg C]

...

Input D2 Temperature [deg C]

...

Input D3 Temperature [deg C]

...

Input D4 Temperature [deg C]

...

Input D5 Temperature [deg C]

Lakeshore Input Names/KPF Location Look-up Tables

(Green) Lakeshore - 192.168.23.181

...

Lakeshore Input

...

KPF Location

...

Notes

...

A

Green Cryostat

Individual Locations TBD

...

B

...

C1

...

C2

...

C3

...

C4

...

C5

...

D1

...

D2

Spectrometer

Individual Locations TBD

...

D3

...

D4

...

D5

(Red) Lakeshore - 192.168.23.182

...

Lakeshore Input

...

KPF Location

...

Notes

...

A

Red Cryostat

Individual Locations TBD

...

B

...

C1

...

C2

...

C3

...

C4

...

C5

...

D1

...

D2

Spectrometer

Individual Locations TBD

...

D3

...

D4

...

D5

Required Alarms

...

Parameter

...

Threshold

...

Notes

...

The temperature sensor cable details differ depending on where the sensors are placed, see “Installation Details and Sensor Locations” table below for more details.

...

Installation Details and Sensor Locations

Description

File

Notes

1

Installation details and sensor locations

Thermal Sensor Locations (Spectrometer and Hallway)

2

Interconnection Diagram
(for Lakeshore sensors within the vacuum chamber)

Image Added

3

Sensor cable harness fabrication details

View file
nameChamber_RTD_Wiring_03.pptx

4

Cable details between Lakeshores and vacuum chamber

Calibration Rack External Cables

...

Relevant Datasheets & Manuals

Description

File

Source

1

Python code that sets up a Lakeshore 224 with KPF settings.

View file
nameLakeshore_224_startup.py.zip

Written by S. Gibson for tests at SSL.

2

Description

File

Source

1

Device Manual

View file
name224_manual.pdf

https://www.lakeshore.com/products/categories/downloads/temperature-products/cryogenic-temperature-monitors/model-224-temperature-monitor

2

Rack Mounting Instructions

View file
namerack-mounting-instructions.pdf

Relevant Software

3

RTD Sensor Datasheet

View file
namesps-siot-temperature-sensors-line-guide-009033-4-en-ciid-45085.pdf

Covers many sensors, the part number of the sensor KPF uses is: HRTS-5760-B-U-1-12.

DigiKey download.

...

Dispatcher Configuration File

Each Lakeshore has a unique name, noted as $(NAME). Currently the names are green and red but that can easily change. That name is used to denote details about the connections.

Each sensor has a unique name, noted as $(sensor). That name is set by the configuration file. The keyword for the sensor is set by the sensor name. The units, data type, and display format are also all defined in the configuration file.

Note, beyond the blocks device and dispatcher each sensor needs its own block. That block named by the sensor name, so, A, B, C1, etc.

Configuration

Value

Explanation

address

192.168.23.181

IP address of the Lakeshore

port

7777

port for connecting to the Lakeshore

poll_time

5

Time in seconds between queries

model

MODEL224

Model number for the device, must in the return string when the model is queried

name

green

Prefix of name, corresponds to which Lakeshore (could be removed)

dispnum

2

Dispatcher number.

name

sensor

Name of the sensor, must be in a block labeled by the internal sensor name (e.g., A, B, C2, etc.)

type

double

The data type of the value

format

%.3f

The ASCII format used for showing the data via KTL. The full data as returned by the device will be stored in the binary format.

unit

deg C

The units of the temperature. The options that actually make sense are deg C and deg K. There is no enforcement of that here but this only impacts the displayed values, it does not change the query made to the device (which can be K or C.)