The Vaisala PTU300 is mounted on the wall chamber side of the thermal enclosureVac Cart. A hose connected to the pressure-measuring spigot on the PTU300 goes through the wall to measure the pressure inside the spectrometer thermal enclosure. This allows pressure within the thermal enclosure to be measured while placing the the heat source of the Vaisala itself outside in the hallway.

...

Description

Qty

Source

Part Number

Notes

Barometer/Humidity Sensor

1

Vaisala

PTU300

Note Vaisala now offers an ethernet interface board for the PTU300, but it has to be installed at the factory.

Serial Interface Cable

1

Vaisala

19446ZZ

Purchased separately and installed by KPF team.

RS-232 to Ethernet Adapter

1

Lantronix

UDS2100

This is the version with two serial ports (but and standard 9-pin RS-232 connectors)

...

Relevant Functional Block Diagram

...

Software Actions

...

Action

...

Notes

...

Read pressure value

...

Value reported as telemetry (see below)

...

Read temperature value

...

Value reported as telemetry (see below)

...

Read relative humidity value

...

Value reported as telemetry (see below)

Required Adjustable Settings

...

Setting

...

When Needed

...

Notes

...

Build

...

Observing

...

Maintenance

...

Pressure measurement frequency

...

Yes

...

No

...

Maybe

...

Set in configuration file. If this needs to be changed, edit configuration file, re-install, and then restart dispatcher.

...

Temperature measurement frequency

...

Yes

...

No

...

Maybe

...

See pressure.

...

Relative humidity measurement frequency

...

Yes

...

No

...

Maybe

...

See pressure.

Required Telemetry

...

Parameter

...

Report Frequency

...

Purpose

...

Notes

...

Pressure measurement

...

Once every 5 sec

...

Environmental monitoring

...

Managed through configuration file

...

Temperature measurement

...

Once every 5 sec

...

Environmental monitoring

...

See pressure.

...

Relative humidity measurement

...

Once every 5 sec

...

Environmental monitoring

...

See pressure.

Required Alarms

The link to the relevant section of the system-wide alarms page is here.

Smaller Footprint Power Supply for Lantronix

1

Tri-Mag, LLC

Tri-Mag PN: L6R12H-120

DigiKey PN: 364-1264-ND

The stock Lantronix power supplies do not work well with the PDUs (as they block two extra outlets). KPF uses these replacement supplies since they only take up one PDU outlet.

...

Relevant Functional Block Diagram

...

Cable Details

Cable

Qty

Details

Notes

Ethernet

1

Cat6 Copper Ethernet Cable M-M, 6 feet long

Entirely within Vac Cart.

...

Pictures of Devices

...

...

Relevant Datasheets & Manuals

...

As part of the KROOT distribution, there is a SerialStream interface which manages connections to and from serial hosts. This includes managing Lantronix connections.

Dispatcher Implementation

Keyword

Value

Type

RETVAL

Value returned by Vaisala.

String. The format is set in the configuration file. Only readable.

TIME

Time stamp of when the last RETVAL was received.

Unix timestamp (binary) or time stamp (ascii). Only readable.

PRES

Pressure value in units of hPa.

Double value from RETVAL, read only.

RELH

Relative humidity in %.

Double value from RETVAL, read only.

TEMP

Temperature in deg C.

Double value from RETVAL, read only.

DISP#STA

Dispatcher status.

Enum and read only. If value is 4 or “Not Connected” the connection to the Vaisala or the Lantonix is not available. If value is 3 or “Connecting” the dispatcher is attempting to connect to the Vaisala through the Lantronix.

DISP#CONN

IP address and port separated by a colon.

Read only string, set by configuration file

DISP#DEV

Device name

Read only string, set by configuration file

Description

File

Source

1

Vaisala PTU300 User’s Guide (2019 version; device purchased in 2014)

View file
namePTU300-Users-Guide-in-English-M210796EN-J.pdf

https://www.vaisala.com/en/products/instruments-sensors-and-other-measurement-devices/instruments-industrial-measurements/ptu300

2

Calibration Certificates

View file
nameVaisala_Calibration_Certificates.pdf

These shipped with our actual KPF device

3

Lantronix UDS 2100 Quick Start Guide

View file
nameUDS2100_QS.pdf

https://www.lantronix.com/products/uds2100/#tab-docs-downloads

34

Lantronix UDS 2100 User Guide

View file
nameUDS2100_UG.pdf

5

Smaller Footprint Power Supply Datasheet

View file
nameL6R12H_Series_2021-01-1.pdf

https://www.lantronixtri-mag.com/products/uds2100/#tab-docs-downlo

Relevant Software

...

Description

...

Files

...

Python 2 software written by a student some years back. These lines of code were used to read values from the Vaisala while it was installed at Keck.

That installation was using a USB interface on a Windows machine. Not sure if the current serial interface will change anything here.

...

View file
nameExample Python 2 Code.zip

wp-content/uploads/2021/05/L6R12H_Series_2021-01.pdf

(Linked off the DigiKey product page)

...

Dispatcher Configuration File

Configuration

Value

Explanation

ip

192.168.23.126

IP address of the Lantronix box.

port

10001

Port that is used for Vaisala, this is port 1.

format_str

form 9.4 "P=" P " " U6 6.4 "T=" T " " U3 6.4 "RH=" RH " " U4 \r \n

Format statement of the output from the Vaisala. At dispatcher start up and at every new connection, this value is sent to the Vaisala. The #.# sets the number of characters (first # ) and the second # sets the number of characters after the decimal point. There must be a space between the characters that notes what the value is (such as P=) and the actual value (the P). The code uses the characters (P=, RH= or T=) to determine what the next value is.

poll_time

5

Time between reads of the Vaisala in seconds.