Introduction
A vent valve is located on the door flange of the main vacuum chamber. It has a solenoid within that is controlled by mains power: the vent valve opens when power is applied, and closes when power is removed.
Within KPF it is remotely-controlled by turning its PDU outlet on or off (via kpfpower).
This valve should only be operated by engineering staff while following the chamber venting procedure. |
...
Relevant Devices
Description | Qty | Source | Part Number |
---|---|---|---|
Vent Valve for NW 25 with 115V Solenoid | 1 | Agilent Varian | L9942302 |
...
Relevant Functional Block Diagram
...
Cable Details
Cable | Qty | DetailsNotes |
---|---|---|
Vent Solenoid Cable (120V) | 1 | NEMA 5-15P cable spliced onto pigtail leads from vent valve. |
...
Relevant Datasheets & Manuals
Description | File | Source | ||||
---|---|---|---|---|---|---|
1 | Relevant Section of Agilent Varian Valve Catalog (with our valve highlighted). | https://www.idealvac.com/files/literature/11_Varian_2011_Vacuum_Valves.pdf | ||||
2 | Full Agilent Varian Valve Catalog |
|