Relevant Devices

Required Adjustable Settings

NOTE: device automatically switches from “LOCAL” to “REMOTE” mode when it receives a remote command through either the USB, RS-232C, or IEEE 488 GPIB interfaces.

...

Setting

...

When Needed

...

Notes

...

Build

...

Observing

...

Maintenance

...

Number of measurements 'n' to include in average.

...

YES

...

NO

...

MAYBE

...

Perhaps start with 10

...

Time delay 't' between measurements used for average.

...

YES

...

NO

...

MAYBE

...

Perhaps start with 0.5 seconds.

...

Device setup parameters follow:

...

Set offset subtraction

...

YES

...

NO

...

MAYBE

...

Equivalent to “ZERO” button on device.

KPF default should be ‘Z0’ (no offset subtraction).

...

Set calibration wavelength

...

YES

...

NO

...

MAYBE

...

Equivalent to wavelength buttons on device.

KPF default value should be 658 nm. Remote command might be 'W0658'.

...

Set attenuator mode

...

YES

...

NO

...

MAYBE

...

“ATTN” button on device.

KPF default should be set to ‘off’ (i.e. no attenuator in place). Remote command is ‘A0’

...

Set measurement units

...

YES

...

NO

...

MAYBE

...

Equivalent to “UNITS” button on device.

KPF default should be Watts (remote command ‘U1').

...

Set averaging mode

...

YES

...

NO

...

MAYBE

...

Equivalent to “AVG” button on device.

KPF default should be set to ‘Fast’ mode (no averaging by device). Remote command for ‘Fast’ mode is ‘F3’.

...

Set audible beep on/off

...

YES

...

NO

...

MAYBE

...

Equivalent to “BEEP” button on device.

KPF default should be ‘off’ (remote command ‘B0’)

...

Set automatic range

...

YES

...

NO

...

MAYBE

...

Equivalent to “AUTO” range on device.

KPF default should be auto-range (remote command ‘R0’).

...

Turn display on/off

...

YES

...

NO

...

MAYBE

...

KPF default should be ‘off’ (remote command ‘K0’). Likely device produces less heat (and isn’t a light source) with the display off.

...

Local lockout

...

YES

...

NO

...

MAYBE

...

I presume KPF default should be ‘locked’ while under remote operation (remote command ‘L1’)

Required Telemetry

...

Parameter

...

Report Frequency

...

Purpose

...

Notes

...

Source 1 intensity

...

TBC

Monitor calibration source outputs over time

...

Report this value if Source 1 is selected at octagon.

...

Source 2 intensity

...

TBC

...

Report this value if Source 2 is selected at octagon.

...

Source 3 intensity

...

TBC

...

Report this value if Source 3 is selected at octagon.

...

Source 4 intensity

...

TBC

...

Report this value if Source 4 is selected at octagon.

...

Source 5 intensity

...

TBC

...

Report this value if Source 5 is selected at octagon.

...

Source 6 intensity

...

TBC

...

Report this value if Source 6 is selected at octagon.

...

Source 7 intensity

...

TBC

...

Report this value if Source 7 is selected at octagon.

...

Source 8 intensity

...

TBC

...

Report this value if Source 8 is selected at octagon.

Data taken with cal unit closed up June 7, 2022.

Wavelengths selected based on central band pass of the two detectors.

Dark was taken with the Solar calibrator unplugged and the intensity monitor out of the beam and the lamps were all off.

Source

Power at 522 nm (Watts)

Power at 735 nm (Watts)

Broad Band

1..72e-05

1.303e-05

UNe Daily

1.442e-05

1.063e-05

ThAr Daily

1.214e-05

9.011e-06

ThAr Gold

1.581e-05

1.153e-05

UNe Gold

2.044e-05

1.508e-05

Dark

7.116e-10

6.056e-10

Broad band on, detector at -15000

1.15e-9

Broad band on, detector at -25000

3.33e-9

Broad band on, detector at -35000

3.77e-8

Description

Qty

Source

Part Number

Lamp Intensity Monitor Controller

1

Newport

1830-R-GPIB

Lamp Intensity Monitor Detector

1

Newport

818-SL/DB

RS-232 to Ethernet Adapter

1

Lantronix

UDS2100

This is the version with two serial ports (and standard 9-pin RS-232 connectors)

Relevant Functional Block Diagram

Most recent version. Going forward, lamp intensity monitor and its connections are not expected to change.

...

Software Actions

...

Action

...

Notes

...

Single measurement.

...

That is, report the Wattage value the detector is currently reporting.

Note calibration unit translation stage has to move detector into beam first.

...

Calculate average of ‘n' discrete single measurements taken ’t' seconds apart.

...

Average several single measurements together to smooth out measurement noise/detector variability.

This value reported as telemetry (see below).

This approach was used for measurements at SSL using a similar device; see example Python code at bottom of this page.

Smaller Footprint Power Supply for Lantronix

1

Tri-Mag, LLC

Tri-Mag PN: L6R12H-120

DigiKey PN: 364-1264-ND

The stock Lantronix power supplies do not work well with the PDUs (as they block two extra outlets). KPF uses these replacement supplies since they only take up one PDU outlet.

...

Relevant Functional Block Diagram

...

Cable Details

Cable

Qty

Details

Notes

Ethernet

1

Cat6 Copper Ethernet Cable M-M, 6 feet long

Entirely within Cal Rack.

DB9 Serial Cable (Straight-Thru)

1

DB9 F-F Straight-thru cable, 1 foot long

Purchased at SFCable.com
(PN D930-01)

DB15 Male-Female Detector Extension Cable

1

BNC F-M Extension Cable, 50-Ohms, 22-Gauge, 12 feet long

McMaster-Carr
(PN 6641T35)

DB15 to BNC Adapter

1

Shipped with the detector and plugs into the DB15 port on the front of the device.

Image Added

...

Pictures of Devices

...

Relevant Datasheets & Manuals

Description

Files

Notes

Example python code from VPH measurement system at SSL. Written for a Newport 1936-C intensity meter, which perhaps uses similar code to the KPF intensity meter (Newport 1830-R-GPIB)

Description

File

Source

1

1830-R-GPIB Quick Start

View file
name1830-R-QUICK-START-GUIDE.pdf

https://www.newport.com/p/1830-R-GPIB

2

1830-R-GPIB Manual

View file
name1830-R-Manual-RevA.pdf

https://www.newport.com/p/1830-R-GPIB

3

Controller Computer Interface Software

View file
namePowerMeter 3.0.4.zip

https://www.newport.com/p/1830-R-GPIB

4

Controller Firmware

View file
name1830-R-Firmware-v1.5.zip

https://www.newport.com/p/1830-R-GPIB

5

818-SL/DB Detector Manual

View file
nameRevB-818-Series-Manual.pdf

https://www.newport.com/p/818-SL--DB

6

818-SL-L Detector Calibration Report

View file
nameNewport_818-SL-L_Calibration_report.pdf

Shipped with actual KPF device

7

Lantronix UDS 2100 Quick Start Guide

View file
nameUDS2100_QS.pdf

https://www.lantronix.com/products/uds2100/#tab-docs-downloads

8

Lantronix UDS 2100 User Guide

View file
nameUDS2100_UG.pdf

https://www.lantronix.com/products/uds2100/#tab-docs-downlo

Relevant Software

9

Smaller Footprint Power Supply Datasheet

View file
name

Example

L6R12H_

intensity_monitor_code.zip

Dispatcher Implementation

Keyword

Value

Type

POW

Power measured by detector every time delay (see above)

Double - read only

AVGPOW

Power measured by detector averaged over a number of steps specified in NAVG

Double - read only

AVGTIME

Time stamp of last average measurement

Time in ascii format or unix timestamp in binary format - read only

NAVG

Number of POW measurements to average to compute AVGPOW. Dispatcher requires > 1.

Integer - writable as as well readable, default set in configuration file

AVGSTEP

Number of POW measurements used in current AVGPOW value. Goes to 0 when NAVG is reached.

Integer - read only.

AVG

Averaging is on or off. Turns off when AVGSTEP == NAVG.

Boolean

SERIALCONN

Connection on or off to the device

Boolean

NEWPORT_ID

ID string returned by device. Only valid when there is a connection.

String - read only.

NEWPORT_REV

Revision of the firmware.

String - read only.

ATT

Attenuator on / off

Enum - 0 Off - 1 On

BEEP

Device Beep on / off

Enum - 0 Off - 1 On

DISPLAY

Front panel display on / off

Enum - 0 Off - 1 On

FILTER

Averaging on the device

Enum - 1 Slow (16) - 2 Medium (4) - 3 Fast (1 or no averaging)

LOCK

Front paneled locked remotely, on locks

Enum - 0 Off - 1 On

RANGE

The device does automatic scaling of the output (0) or has eight preset ranges

Enum - 0 Auto - 1 2 3 4 5 6 7 8

UNITS

The output units of the power measurement. Can be relative (REL) to a previous value.

Enum - 1 Watts - 2 dB - 3 dBm - 4 REL

ZERO

Use a previous value as a zero point, can be turned on or off.

Enum - 0 Off - 1 On

WAVE

Wavelength of light used for the detector calibration, specified in nanometers. This is only actually useful if you know the wavelength of the source and it is reasonably monochromatic.

Integer - dispatcher enforces range between 400 and 1100

DISP#STA

Dispatcher status.

Enum and read only. If value is 4 or “Not Connected” the connection to the Vaisala or the Lantonix is not available. If value is 3 or “Connecting” the dispatcher is attempting to connect to the Vaisala through the Lantronix.

DISP#CONN

Dispatcher connection details, contains IP address and port separated by a colon

String

Series_2021-01-1.pdf

https://www.tri-mag.com/wp-content/uploads/2021/05/L6R12H_Series_2021-01.pdf

(Linked off the DigiKey product page)

...

Dispatcher Configuration File Details

Configuration

Value

Explanation

dipsnum

1

Dispatcher number

ip

192.168.23.178

IP address of the Lantronix

port

10001

Port that is used, 10001 means port 1

Lantronix serial config

19200, 8N1, No flow control

Serial port settings, matched to Newport 1830-R setting.

poll_time

0.5

Time between data requests for the measured power value.

navg

10

Default starting value for NAVG. Keyword can be written.

wave

658

Default starting value for the wave parameter, the value is nm.

units

1

Default starting value for units, this is Watts.

filter

3

Default starting value for the filter, this is the Fast value.

display

0

Default starting value for the display, this is the off value.

...

Calibration Intensity Measurement

Software Actions

The calibration intensity monitor measures the intensity of calibration lamps when it is in the beam. This dispatcher automates some of the configuration and writes the measured power to a keyword whose name matches the calibration source when the measurement being performed.

The measurement only occurs when the AVG value of the calibration intensity monitor is On. It also checks that the lamp is on for which ever source is selected by the octagon stage, and that the monitor is in the beam. If either state changes during a measurement, the keyword is no longer updated. Currently it does not check the calibration intensity monitor itself.

Dispatcher Implementation

Keyword

Value

Type

MEASURING

Yes / No - Is the average power being recorded.

boolean

INPUT

Octagon position selected, that is also the name of the keyword that the power will be recorded to.

enumerated

<LAMP_NAME>

The last average power value recorded.

Double

<LAMP_NAME>_TIME

The time that the last power value was recorded.

Double

...

Data taken with cal unit closed up June 7, 2022.

Wavelengths selected based on central band pass of the two detectors.

Dark was taken with the Solar calibrator unplugged and the intensity monitor out of the beam and the lamps were all off.

Source

Power at 522 nm (Watts)

Power at 735 nm (Watts)

Broad Band

1..72e-05

1.303e-05

UNe Daily

1.442e-05

1.063e-05

ThAr Daily

1.214e-05

9.011e-06

ThAr Gold

1.581e-05

1.153e-05

UNe Gold

2.044e-05

1.508e-05

Dark

7.116e-10

6.056e-10

Broad band on, detector at -15000

1.15e-9

Broad band on, detector at -25000

3.33e-9

Broad band on, detector at -35000

3.77e-8