Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Description

Qty

Source

Part Number

Notes

Shutter Controller

1

Uniblitz

VED24

Shutter

1

Uniblitz

NS65B1T0-VED

65mm Optical Shutter

Bi-stable (only requires power for transitions)

Ordering Options:

Note: as of 2021 the electronic sync option was not available for this shutter. Uniblitz said that option was only possible for this shutter as part of a custom, large-quantity order.

...

Relevant Functional Block Diagram

...

Cable

Qty

Details

Sci/Sky Scrambler Shutter (BNC)

1

See Aux Rack External Cables

Uniblitz Custom-Length 510F Cable
(30-feet Long)

1

View file
nameQuote # 071018SS4.pdf

...

Photos of Devices

...

Installation Instructions

The settings below must only be used for the NS65B shutter, or damage will result.

The shutter controller has a set of DIN “Function Switches” (second from right in image below) and an Octal switch for “Pulse Duration” (far right in image below) that need to be configured for the particular shutter being used:

...

The function switch has the following settings:

...

For the DSS25B NS65B shutters the function switch and pulse duration settings are:

...

Note the setting for Switch 1 does not depend on the shutter type, but should be set to ‘UP’ for ‘LOCAL’ (this disables the USB port and enables the BNC Pulse Input):

Note also that there is also a toggle switch on the front that sets the ‘Normally Open’ or ‘Normally Closed’ polarity of the switch (second from left in image above). The manual is not clear on how this affects a bi-stable shutter. For consistency within KPF, all shutter controllers should be set to ‘N.O.’. Each KPF VED has a label on the front to indicate this, as show in the picture above.

...