General Principles
DO NOT use
control-c
to stop a script, this can leave the instrument in an unsafe state. Usekpfconfig.SCRIPTSTOP
to request a stop (also available as a button in the KPF OB GUI).
...
Setting SCRIPTSTOP to “Yes” will request that the running script terminate in an orderly fashion. This may take several minutes depending on when the running script reaches a sensible breakpoint. It is important to use SCRIPTSTOP to halt a script (instead of hitting ctrl-c
) because cleanup actions will be performed after a SCRIPTSTOP (e.g. turning off laps to preserve their lifetime).
...
The agitator can be reenabled by simply setting the keyword to “Yes”. This should only be done by WMKO staff and should only be done if the agitator is fully functional. A broken or misbehaving agitator mechanism presents a significant danger to the science fibers.
...
The cal source has likely been disabled for a reason. Reenabling it should only be done by WMKO staff with knowledge of why it was disabled in the first place.
SlewCal or Simultaneous Calibration Source is Wrong
...
These two sources are set from the same place and are always kept to the same value using the kpfconfig.SIMULCALSOURCE
keyword.
Solution
This value should only be changed by WMKO staff. The choice of calibration source is influenced by our desire to maintain the lifetime of the various lamps and calibration sources and by the need for warmup times on certain sources. This is not something the user should adjust, contact a WMKO Staff Astronomer if you feel it is set incorrectly.
...
The detector has likely been disabled for a reason. Reenabling it should only be done by WMKO staff with knowledge of why it was disabled in the first place.
Detector Fails to Trigger Without Error
...
Green or Red CCDPOWER is “Not Configured”
Symptom
The CCDPOWER keyword in the kpfgreen or kpfred service is “Not Configured”. This may manifest as the system not being able to take exposures.
Problem
The core problem is not yet clear, but this appears to be something in the Archon controller itself. Investigation is ongoing as of this writing. “It is as if the Archon glitched and lost its ACF.”
Solution
Restart the relevant kpfgreen or kpfred service (we will use kpfgreen in this example):
kpf restart kpfgreen
After the restart, the heater configuration must immediately be reset to the nominal values to restore temperature control, this will also turn on CCDPOWER:
kpfmon-heater-copy green copytolive
Ca HK Detector Stuck
Symptom
The HK Note: The above solution was made obsolete when the Archons were removed from the heater control loops.
Green or Red CCDPOWER Fails to turn on
Symptom
When trying to power up the detector by setting the CCDPOWER
keyword to “On”, the keyword does not successfully transition to on. The stderr file for the relevant camerad process (e.g. `gcamerad.stderr`) contains statements like:(Archon::Interface::archon_cmd) ERROR: Archon controller returned error processing command: POWERON
Problem
There is something wrong in the ACF file for the Archon.
Solution
Resend the ACF file by either modifying the appropriate ACF
keyword or using the kpfSetReadModeNormal
or kpfSetReadModeFast
scripts.
Ca HK Detector Stuck
Symptom
The HK detector is stuck in the exposing state.
...
This connects to the Raspberry Pi controller in the dome enclosure. The username and IP address has configured in the ~/.ssh/config
file on kpfserver (you can ssh manually using pi@192.168.23.244
) and the SSH key for kpfserver has been installed on the Pi, so it should not ask for a password, but if it does, the password in in the usual showpasswords
location.View the dome log file in the ~/dome.log
and look for errors which indicate the nature of the problem location.
View the dome log file in the ~/dome.log
and look for errors which indicate the nature of the problem.
The ~/grep_for_dome_error
script will exclude many of the noisy, not useful lines in the dome.log
file and help with examining the log.
Solution
If the log file indicates overcurrent on the motors is the issue. Ensure the mechanisms are clear of obstruction and reasonably well balanced (it doesn’t need to be perfect).
...
Code Block |
---|
2024-07-09 19:09:37,752 WARNING Operation timed out (35.0 secs), max measured motor
current: 0.0 A. |
The ~/grep_for_dome_error
script will exclude many of the noisy, not useful lines in the dome.log
file and help with examining the log.
Solution
Reboot the controller (raspberry pi): sudo reboot
and restart the kpfsocal3 dispatcher: kpf restart kpfsocal3
.