Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Table of Contents:

Table of Contents

...

Introduction

Two vacuum gauges are located on the main KPF vacuum chamber and are monitored by a Leybold Graphix Two gauge controller.

...

Relevant Devices

Description

Qty

Source

Part Number

Notes

Vacuum Gauge Controller

1

Leybold

GRAPHIX

Graphix Two

Controls the gauges on the chamber

High Vac gauge

1

Leybold

PTR 90 N

Combo Pirani and Cold Cathode gauge

Low Vac gauge

1

Leybold

TTR 91 N S

Pirani gauge

RS-232 to Ethernet Adapter

1

Lantronix

UDS 2100

This is the version with two serial ports

Functional Block Diagram

...

(This device at lower right)

Software Actions

...

Action

...

Notes

...

Read Channel 1 Pressure Value

...

Value reported as telemetry (see below)

...

Read Channel 2 Pressure Value

...

Value reported as telemetry (see below)

...

Channel 1 Gauge Status

...

Value reported as telemetry (see below)

...

Channel 2 Gauge Status

...

Value reported as telemetry (see below)

...

Set Point 1 Status

...

Value reported as telemetry (see below)

Required Adjustable Settings

...

Setting

...

When Needed

...

Notes

...

Build

...

Observing

...

Maintenance

...

Telemetry Read Cadence

...

Yes

...

No

...

Maybe

Device Set-up Parameters

View file
nameGRAPHIX Two Software Settings - v3.xlsx

Required Telemetry

...

Parameter

...

Report Frequency

...

Purpose

...

Notes

...

Channel 1 Pressure Value

...

Once per minute [TBD]

...

Channel 2 Pressure Value

...

Once per minute [TBD]

...

Channel 1 Gauge Status

...

Once per minute [TBD]

...

Channel 2 Gauge Status

...

Once per minute [TBD]

...

Set Point 1 Status

...

Once per minute [TBD]

...

Check on gate valve position

Required Alarms

Parameter

Threshold

Notes

1

Chamber pressure too low

TBD

2

(but standard 9-pin RS-232 connectors)

Smaller Footprint Power Supply for Lantronix

1

Tri-Mag, LLC

Tri-Mag PN: L6R12H-120

DigiKey PN: 364-1264-ND

The stock Lantronix power supplies do not work well with the PDUs (as they block two extra outlets). KPF chose these replacement supplies since they only take up one PDU outlet.

...

Functional Block Diagram

...

Cable Details

Cable

Qty

Details

Notes

Ethernet

1

Cat6 Copper Ethernet Cable M-M, 6 feet long

Entirely within Vac Cart.

RS-232

1

DB9, M-F, Straight through, 3 feet long

Entirely within Vac Cart.

Gauge Cables (Vac Cart)

2

Shielded ethernet cables, RJ-45, M-M, 6 feet long

Entirely within Vac Cart.

Gauge Cables (Chamber)

2

Shielded ethernet cables, RJ-45, M-M, 10 feet long

Bulkhead Coupler (Dual)

1

Universal Sub-Panel Black, 2 Black Feed-Thru Couplers, RJ45K (8x8K) Straight
L-Com PN: USP2ECFB8SKB

...

Pictures of Devices

...

Relevant Datasheets & Manuals

Description

File

Source

12

Description

File

Source

1

Leybold Graphix Two Manual

View file
name300550402_002_C1_GRAPHIX_123_EN.pdf

https://www.leyboldproducts.com/media/pdf/0f/9e/dc/300550402_002_C1_GRAPHIX_123_EN.pdf

2

Relevant Software

‘Vacuum Measurement’ brochure containing specs on various gauges, including both of our gauges (PTR 90 N and TTR 91 N S)

View file
name3610019902_Sensors_2020_web.pdf

https://www.leyboldproducts.us/products/measuring-and-analysis-instruments/pressure-measurement/active-sensors/penningvac-transmitters/ptr-90-n/1542/penningvac-ptr-90-n

3

PTR 90 N Vacuum Gauge ‘Technical Data’ sheet

View file
namePTR_90N_TD_230072V02_EN57d6f9450e7ae.pdf

https://www.leyboldproducts.us/products/measuring-and-analysis-instruments/pressure-measurement/active-sensors/penningvac-transmitters/ptr-90-n/1542/penningvac-ptr-90-n

4

CAD Model for PTR 90 N gauge.

View file
nameCAD_230072V02-STP.zip

https://www.leyboldproducts.us/products/measuring-and-analysis-instruments/pressure-measurement/active-sensors/penningvac-transmitters/ptr-90-n/1542/penningvac-ptr-90-n

5

Datasheet for TTR 91 N S gauge.

Seems a bit outdated, and was downloaded from a vac supply company (not Leybold). However there was nothing on this part number at Leybold (closest was a TTR RNS model, which might be the newer version of our gauge).

View file
namevacshop-THERMOVAC_New-TTR91N.pdf

http://vacshop.ir/wp-content/uploads/2016/02/vacshop-THERMOVAC_New-TTR91N.pdf

6

Lantronix UDS 2100 Quick Start Guide

View file
nameUDS2100_QS.pdf

https://www.lantronix.com/products/uds2100/#tab-docs-downloads

7

Lantronix UDS 2100 User Guide

View file
nameUDS2100_UG.pdf

8

Smaller Footprint Power Supply Datasheet

View file
nameL6R12H_Series_2021-01.pdf

https://www.tri-mag.com/wp-content/uploads/2021/05/L6R12H_Series_2021-01.pdf

(Linked off the DigiKey product page)