Introduction

Given the high stability requirements of KPF, the main spectrometer sits within a vacuum chamber in order to isolate the spectrometer from its environment. This provides isolation against changes in atmospheric pressure and also convective thermal disturbances.

The vacuum pump system for the KPF vacuum chamber consists of a turbo pump backed by a roots pump. The Leybold TurboVac pump is controlled via a Leybold Turbo.Control controller, which contains a webserver to allow control of the pump via a webpage in a browser. The Leybold EcoDry roots pump is simply turned on and off via a PDU outlet on the Vac Cart.

Within KPF the turbo and roots pump run continuously.

The Turbo.Control controller also allows connection of two vacuum gauges; we use both of these ports to connect a low and a high vacuum gauge. These two gauges monitor the vacuum pressure within the pipe going to the main KPF vacuum chamber. They are mounted on the Vac Cart and sit on the Vac Cart side of the main chamber gate valve.

Direct computer control of the TurboVac pump is possible via serial commands, however creating a full KTL dispatcher for the pump was deemed too difficult given the dozens of commands required, as well as the fail-safes required to not damage the pump. As such we decided to use the vendor-supplied web-based software to control the pump, which has all the features we need built-in. Since this webpage also provides pump telemetry, we wrote a KPF KTL dispatcher to monitor this webpage and write telemetry values to KTL keywords.

...

Relevant Devices

Description

Qty

Source

Part Number

Notes

Turbo Pump

1

Leybold

LH TurboVac 450iX

Turbo Pump Controller

1

Leybold

Turbo.Control i

Roots Pump

1

Leybold

LH EcoDry65

Slaved to Turbo Pump (no direct control)

No computer control

High Vac gauge

2

1

Leybold

PTR 90 N

Combo Pirani and Cold Cathode gauge

Low Vac gauge

2

1

Leybold

TTR 91 N S

Pirani gauge

...

Relevant Functional Block Diagram

...

This page is focused on the devices in yellow boxes.

Software Actions

...

Action

...

Notes

...

Read Telemetry

...

Pump and gauge telemetry available through Pump Controller webserver (Leybold turbo.control)

...

Required Adjustable Settings

...

Setting

...

When Needed

...

Notes

...

Build

...

Observing

...

Maintenance

...

N/A

...

Telemetry only (no control)

Required Telemetry

...

Parameter

...

Report Frequency

...

Purpose

...

Notes

...

Actual Power

Turbo pump health monitoring

Given on webserver status page (and “Data Viewer” page)

...

Actual Frequency

...

Bearing Temperature

...

Actual Voltage

...

Gauge 1 Pressure

...

Vac cart pipe pressure

...

Gauge 2 Pressure

Required Alarms

...

Parameter

...

Threshold

...

Notes

...

...

Cable Details

Cable

Qty

Details

Notes

Ethernet

1

Cat6 Copper Ethernet Cable M-M, 6 feet long

Entirely within Vac Cart.

Gauge Cable

2

Shielded RJ-45

Shipped with Gauges

RS-485

1

DB-9 M-M, 3 feet long

Shipped with TurboControl or TurboPump

...

Photos of Devices

Image Added

...

Relevant Datasheets & Manuals

Description

File

Source

1

EcoDry Roots Pump Manual

View file
nameEcoDry_Manual.pdf

2

Turbo Pump Operating Manual

View file
name300554863_002_C0_TURBOVAC_i_X_EN.pdf

3

Turbo.control Gauge and Pump Controller

View file
name300702826_002_C0_TurboControl_i_Software.pdf

3

4

Serial Interfaces for TURBOVAC

View file
name300450826_002_C0_Serial-TiX.PDF

4

5

Pump Power Supply “Brief Instructions”

View file
nameTurbo_Power_Supply_Manual.pdf

5

Relay Box manual

(Allows Turbo Pump to control Roots Pump

6

‘Vacuum Measurement’ brochure containing specs on various gauges, including both of our gauges (PTR 90 N and TTR 91 N S)

View file
name

Relay

3610019902_Sensors_

Box

2020_

Manual

Description

File

Notes

1

Vac pump built-in web server

View file
nameVacuum pump webserver.pdf

Offers control and (potentially) telemetry

2

web.pdf

Relevant Software

https://www.leyboldproducts.us/products/measuring-and-analysis-instruments/pressure-measurement/active-sensors/penningvac-transmitters/ptr-90-n/1542/penningvac-ptr-90-n

7

PTR 90 N Vacuum Gauge ‘Technical Data’ sheet

View file
namePTR_90N_TD_230072V02_EN57d6f9450e7ae.pdf

https://www.leyboldproducts.us/products/measuring-and-analysis-instruments/pressure-measurement/active-sensors/penningvac-transmitters/ptr-90-n/1542/penningvac-ptr-90-n

8

CAD Model for PTR 90 N gauge.

View file
nameCAD_230072V02-STP.zip

https://www.leyboldproducts.us/products/measuring-and-analysis-instruments/pressure-measurement/active-sensors/penningvac-transmitters/ptr-90-n/1542/penningvac-ptr-90-n

9

Datasheet for TTR 91 N S gauge.

Seems a bit outdated, and was downloaded from a vac supply company (not Leybold). However there was nothing on this part number at Leybold (closest was a TTR RNS model, which might be the newer version of our gauge).

View file
namevacshop-THERMOVAC_New-TTR91N.pdf