Table of Contents:


Page Status

Date

Initials

Change


Relevant Devices

Description

Qty

Source

Part Number

Green CCD Cold Head
Temperature Controller

1

Instec

mK2000

Red CCD Cold Head
Temperature Controller

1

Instec

mK2000

Green LN2 Pump

1

Instec

LN2-SYS

Red LN2 Pump

1

Instec

LN2-SYS


Relevant Functional Block Diagram


Software Actions

Action

Notes

1

Read telemetry (Green)

Continuously running

2

Read telemetry (Red)

3

Upload PID parameter table (Green)

Engineering and maintenance only

4

Upload PID parameter table (Red)

5

Cool-down mode (Green)

Engineering and maintenance only

6

Cool-down mode (Red)

7

Warm-up mode (Green)

Engineering and maintenance only

8

Warm-up mode (Red)


Operational Modes

  • Link to Instec tab within Operational Modes Spreadsheet

    • No modes yet for controller (always on device??)


Required Adjustable Settings

Setting

When Needed

Notes

Build

Observing

Maintenance

PID Parameters

YES

NO

MAYBE


Required Telemetry

Parameter

Report Frequency

Purpose

Notes

1

Green Channel Instec

2

Device Serial Number?

At startup

Include in log just in case device is swapped out in future

*IDN?

(third field of reply)

3

Device Rev Number

At startup

Include in log just in case device is swapped out in future

*IDN?

(fourth field of reply)

4

Cold Head Temperature Value

1 Hz

Monitor CCD temperature

TEMP:CTEM?
5

Heating/cooling enabled status

1 Hz [TBD]

TEMP:CHSW?
6

System status

TEMP:STAT?
7

Temperature set point

TEMP:SPO?
8

Power value

TEMP:POW?
9

Red Channel Instec

10

Device Serial Number?

At startup

Include in log just in case device is swapped out in future

*IDN?

(third field of reply)

11

Device Rev Number

At startup

Include in log just in case device is swapped out in future

*IDN?

(fourth field of reply)

12

Cold Head Temperature Value

1 Hz

Monitor CCD temperature

TEMP:CTEM?
13

Heating/cooling enabled status

1 Hz [TBD]

TEMP:CHSW?
14

System status

TEMP:STAT?
15

Temperature set point

TEMP:SPO?
16

Power value

TEMP:POW?

Required Alarms

Parameter

Threshold

Notes

1

2


Relevant Datasheets & Manuals

Description

File

Source

1

mK2000 User Manual

email from Instec

2

mK2000 Command Reference

email from Instec


Relevant Software

Description

File

Source

1

Instec said: “While we do have sample code for Python, it is based on a USB connection, not Ethernet. 

Our documentation on Ethernet programming is limited because it isn’t very frequently used for custom applications.

This ZIP file contains example code for python using a USB connection. It could serve as the basis for an ethernet version.”

email from Instec

2

Ethernet communication setup from Instec.

Mainly describes how to write code to find Instec devices on a network.

Instec said: “

email from Instec

3

4

Instec App for Windows
(standalone program; not useful for KTL but included here for completeness)

email from Instec