Relevant Devices

Description

Qty

Source

Part Number

Lamp Intensity Monitor Controller

1

Newport

1830-R-GPIB

Lamp Intensity Monitor Detector

1

Newport

818-SL/DB

RS-232 to Ethernet Adapter

1

Lantronix

UDS2100

This is the version with two serial ports (and standard 9-pin RS-232 connectors)


Relevant Functional Block Diagram


Software Actions

Action

Notes

1

Read light intensity value.

Note calibration unit translation stage has to move detector into beam first.
Value reported as telemetry (see below)

2


Required Adjustable Settings

NOTE: device automatically switches from “LOCAL” to “REMOTE” mode when it receives a remote command through either the USB, RS-232C, or IEEE 488 GPIB interfaces.

Setting

When Needed

Notes

Build

Observing

Maintenance

Set offset subtraction

Equivalent to “ZERO” button on device.

Default should be ‘Z0’ (no offset subtraction).

Set calibration wavelength

Equivalent to wavelength buttons on device.

Default value should be 658 nm. Remote command might be 'W0658'.

Set attenuator mode

“ATTN” button on device.

Default should be set to ‘off’ (i.e. no attenuator in place). Remote command is ‘A0’

Set measurement units

Equivalent to “UNITS” button on device.

Default should be Watts (remote command ‘U1').

Set averaging mode

Equivalent to “AVG” button on device.

Default should be set to ‘Fast’ mode (no averaging by device). Remote command for ‘Fast’ mode is ‘F3’.

Set audible beep on/off

Equivalent to “BEEP” button on device.

Default should be ‘off’ (remote command ‘B0’)

Set automatic range

Equivalent to “AUTO” range on device.

Default should be auto-range (remote command R0).

Turn display on/off

Default should be ‘off’ (remote command ‘K0’).

Local lockout

Default should be ‘locked’ while under remote operation (remote command ‘L1’)


Required Telemetry

Parameter

Report Frequency

Purpose

Notes

1

Source 1 intensity

TBC

Monitor calibration source outputs over time

Report this value if Source 1 is selected at octagon.

2

Source 2 intensity

TBC

Report this value if Source 2 is selected at octagon.

3

Source 3 intensity

TBC

Report this value if Source 3 is selected at octagon.

4

Source 4 intensity

TBC

Report this value if Source 4 is selected at octagon.

5

Source 5 intensity

TBC

Report this value if Source 5 is selected at octagon.

6

Source 6 intensity

TBC

Report this value if Source 6 is selected at octagon.

7

Source 7 intensity

TBC

Report this value if Source 7 is selected at octagon.

8

Source 8 intensity

TBC

Report this value if Source 8 is selected at octagon.


Required Alarms

Parameter

Threshold

Notes

1

Calibration source intensity low

TBD

2


Relevant Datasheets & Manuals

Description

File

Source

1

1830-R-GPIB Quick Start

https://www.newport.com/p/1830-R-GPIB

2

1830-R-GPIB Manual

https://www.newport.com/p/1830-R-GPIB

3

Controller Computer Interface Software

https://www.newport.com/p/1830-R-GPIB

4

Controller Firmware

https://www.newport.com/p/1830-R-GPIB

5

818-SL/DB Detector Manual

https://www.newport.com/p/818-SL--DB

6

Lantronix UDS 2100 Quick Start Guide

https://www.lantronix.com/products/uds2100/#tab-docs-downloads

7

Lantronix UDS 2100 User Guide

https://www.lantronix.com/products/uds2100/#tab-docs-downlo

Relevant Software

Description

Files

Notes

Example python code from VPH measurement system at SSL. Written for a Newport 1936-C intensity meter, which perhaps uses similar code to the KPF intensity meter (Newport 1830-R-GPIB)