Table of Contents:
Introduction
Relevant Devices
Description | Qty | Source | Part Number |
---|---|---|---|
Andor Newton 920P-BU | |||
Relevant Functional Block Diagram
Software Actions
Action | SDK Command | Purpose | Notes | |
---|---|---|---|---|
1 | Turn Cooler On | CoolerON | Turns cooler on. | Needed for engineering. |
2 | Turn Cooler Off | CoolerOFF | Turns cooler off. | Needed for engineering. |
3 |
Required Adjustable Settings
Setting | When Needed | Notes | ||
---|---|---|---|---|
Build | Observing | Maintenance | ||
SetBaselineOffset | Sets Bias level for CCD. Engineering-level change. |
Required Telemetry
Parameters for image headers:
Parameter (SDK Command) | Purpose | Notes | |
---|---|---|---|
1 | GetCameraSerialNumber | Logs which hardware was used for the image. Mainly useful if we have to replace the detector at some point (keeps track of which detector was used when) | |
2 | GetCurrentPreAmpGain | ||
3 | GetDetector | ||
4 | GetHardwareVersion | ||
5 | GetHeadModel | ||
6 | GetPixelSize | ||
7 | GetSoftwareVersion | ||
8 | GetUSBDeviceDetails | ||
9 | GetVersionInfo |
Parameters for KPF telemetry database:
Parameter | SDK Command | Report Frequency | Purpose | Notes | |
---|---|---|---|---|---|
1 | CCD Temperature | GetTemperature | Monitor CCD temperature | GetTemperaturePrecision command likely useful to find available precision from our Newton. | |
2 | |||||
3 | |||||
4 | |||||
5 | |||||
6 |
Start-up Parameters
Parameter (SDK Command) | Purpose | Notes | |
---|---|---|---|
1 | SetBaselineClamp | Enabling baseline clamp ensures that the bias level does not change between images. | |
2 | |||
3 | |||
4 | |||
5 | |||
6 |
Ashley’s settings (to fold into tables above):
Horizontal Clock Speed: 0.5MHz
Vertical Clock Speed: 51.3 microseconds
Preamp Gain Setting: 2
Baseline Clamp: On
Subframe: We will end up taking a subframe in the horizontal 1024 pixel direction (not sure of the bounds currently)
Binning: 1x1, We will never bin in standard mode (but it wouldn’t hurt to have the option in case some unique user comes along)
Cooling: tbd on default
Required Alarms
Parameter | Threshold | Notes | |
---|---|---|---|
1 | |||
2 |
Relevant Datasheets & Manuals
Description | File | Source | |
---|---|---|---|
1 | Andor Newton Manual | Email from Andor sales rep (Mike Zemlan) | |
2 |
Relevant Software
Description | File | Source | |
---|---|---|---|
1 | Capability list for our Newton (terse and verbose versions) | Will Deich’s query program | |
2 | Andor SDK 2 Manual Note SDK 2 is what we need as it’s for CCD detectors (SDK 3 is for CMOS detectors). | Direct link for these file sent by Andor sales rep (Mike Zemlan) | |
3 | Andor SDK 2 (for CCDs) for Linux | ||
4 | Andor SDK 2 (for CCDs) for Windows |
0 Comments