Table of Contents:
Relevant Devices
Description | Qty | Source | Part Number | |
---|---|---|---|---|
Lamp Intensity Monitor Controller | 1 | Newport | 1830-R-GPIB | |
Lamp Intensity Monitor Detector | 1 | Newport | 818-SL/DB | |
RS-232 to Ethernet Adapter | 1 | Lantronix | UDS2100 | This is the version with two serial ports (and standard 9-pin RS-232 connectors) |
Relevant Functional Block Diagram
Software Actions
Action | Notes | |
---|---|---|
1 | Read light intensity value. | Note calibration unit translation stage has to move detector into beam first. |
2 |
Required Adjustable Settings
NOTE: device automatically switches from “LOCAL” to “REMOTE” mode when it receives a remote command through either the USB, RS-232C, or IEEE 488 GPIB interfaces.
Setting | When Needed | Notes | ||
---|---|---|---|---|
Build | Observing | Maintenance | ||
Set offset subtraction | Equivalent to “ZERO” button on device. Default should be ‘Z0’ (no offset subtraction). | |||
Set calibration wavelength | Equivalent to wavelength buttons on device. Default value should be 658 nm. Remote command might be 'W0658'. | |||
Set attenuator mode | “ATTN” button on device. Default should be set to ‘off’ (i.e. no attenuator in place). Remote command is ‘A0’ | |||
Set measurement units | Equivalent to “UNITS” button on device. Default should be Watts (remote command ‘U1'). | |||
Set averaging mode | Equivalent to “AVG” button on device. Default should be set to ‘Fast’ mode (no averaging by device). Remote command for ‘Fast’ mode is ‘F3’. | |||
Set audible beep on/off | Equivalent to “BEEP” button on device. Default should be ‘off’ (remote command ‘B0’) | |||
Set automatic range | Equivalent to “AUTO” range on device. Default should be auto-range (remote command R0). | |||
Turn display on/off | Default should be ‘off’ (remote command ‘K0’). | |||
Local lockout | Default should be ‘locked’ while under remote operation (remote command ‘L1’) |
Required Telemetry
Parameter | Report Frequency | Purpose | Notes | |
---|---|---|---|---|
1 | Source 1 intensity | TBC | Monitor calibration source outputs over time | Report this value if Source 1 is selected at octagon. |
2 | Source 2 intensity | TBC | Report this value if Source 2 is selected at octagon. | |
3 | Source 3 intensity | TBC | Report this value if Source 3 is selected at octagon. | |
4 | Source 4 intensity | TBC | Report this value if Source 4 is selected at octagon. | |
5 | Source 5 intensity | TBC | Report this value if Source 5 is selected at octagon. | |
6 | Source 6 intensity | TBC | Report this value if Source 6 is selected at octagon. | |
7 | Source 7 intensity | TBC | Report this value if Source 7 is selected at octagon. | |
8 | Source 8 intensity | TBC | Report this value if Source 8 is selected at octagon. |
Required Alarms
Parameter | Threshold | Notes | |
---|---|---|---|
1 | Calibration source intensity low | TBD | |
2 |
Relevant Datasheets & Manuals
Description | File | Source | |
---|---|---|---|
1 | 1830-R-GPIB Quick Start | ||
2 | 1830-R-GPIB Manual | ||
3 | Controller Computer Interface Software | ||
4 | Controller Firmware | ||
5 | 818-SL/DB Detector Manual | ||
6 | Lantronix UDS 2100 Quick Start Guide | https://www.lantronix.com/products/uds2100/#tab-docs-downloads | |
7 | Lantronix UDS 2100 User Guide |
Relevant Software
Description | Files | Notes |
---|---|---|
Example python code from VPH measurement system at SSL. Written for a Newport 1936-C intensity meter, which perhaps uses similar code to the KPF intensity meter (Newport 1830-R-GPIB) |
0 Comments