Table of Contents:


System Overview

The CCD thermal control system…

See below for details on the sensors and heaters (including KTL keyword names for each).


Sensor Details (and KTL Keywords)

Cryostat Thermal Sensor Locations

Sensor locations and KTL keywords are shown below.

Lakeshore Channel Mapping (Cryostat Sensors Only)

See Thermal Sensor Locations (Spectrometer and Hallway) for entire list of Lakeshore sensors.

The connectors on the various Lakeshore cable harnesses are labelled 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12.

This diagram indicates how they connect to the Lakeshore rear panel:

Green Lakeshore (cryostat sensors only)

Connector Number

Lakeshore Channel

KTL Keyword

1

Green Lakeshore D2

CRYOSTAT_G1

Within cryostat; likely on detector mounting plate to one side of CCD (opposite side to CRYOSTAT_G3)

2

Green Lakeshore D3 (not functioning)

CRYOSTAT_G2

Within cryostat; likely at the detector end of the cold strap

3

Green Lakeshore D4

CRYOSTAT_G3

Within cryostat; likely on detector mounting plate to one side of CCD (opposite side to CRYOSTAT_G1)

Red Lakeshore (cryostat sensors only)

Connector Number

Lakeshore Channel

KTL Keyword

1

Red Lakeshore D2

CRYOSTAT_R1

Within cryostat; likely on detector mounting plate to one side of CCD (opposite side to CRYOSTAT_R3)

2

Red Lakeshore D3 (not functioning)

CRYOSTAT_R2

Within cryostat; likely at the detector end of the cold strap

3

Red Lakeshore D4

CRYOSTAT_R3

Within cryostat; likely on detector mounting plate to one side of CCD (opposite side to CRYOSTAT_R1)