This is a page to track lenslet mask performance issues and the steps we take to troubleshoot. So far the lenslet mask performance has become more unreliable. Due to the lack of encoders, there is no feedback to track which lenslet mask position is really in use. The unreliable behavior has made it more difficult and time consuming to make rectification matrices.

Default lenslet mechanism setting:

A few things to try to troubleshoo

Perhaps the flat lamp has died, saw no light at all in scans taken after 6/6. Flux in scans taken on 6/6 was orders of magnitude fainter than what it should be.
If the flat lamp was failing, it could explain spotty illumination in previous scans. The lamp brightness would have become unstable.

There were RPC errors,  lenslet stepping errors, and lenslet stage motor update errors.

ofls keyword service is moved to osirisbuild.
On napili,
[760] osrseng@napili: source ~/.cshrc

[761] osrseng@napili: lamp var

                      variable = 0

Problematic recmat scan symptoms: