2023-05-10 at 15:30 HST: KPF Red Detector deltaT=+0.4/-0.2 C
During troubleshooting of green, we twice set kpfred.CF_BASE_ENA to on as instructed by EMIR notices, but off is apparently the nominal state. This caused temperature excursion, but small as it was caught early.
Grafana plot: http://vm-dashboards.keck.hawaii.edu/d/VoSaQfx4l/kpf-red-thermal-control?orgId=1&from=1683763200000&to=1683784800000