NIRSPEC Tips and Tricks
Some tips and tricks for NIRSPEC
Telescope GUI
ssh -X k2obstcs@vm-k2obs
tcsgui -k2
SCAM and SPEC
Different commands are addressed with a sufix:
2 for scam and nothing for spec. Examples:
cdata
for spec
cdata2
for scam
Exposure Time definitions
3 types of script for setting exp times: itime, tint, and atint
itime
sets the keyword itime. The keyword has units of milliseconds, the script itime has units of seconds
tint
sets the exp time in seconds
atint
defines the exp time in an efficient way always using MCDS. It gets stuck in sampmode=3. For instance atint2 1.5 will set sampmode to 3 and increase the exp time to the minimum. To revert to simple CDS use sampmode2 2
. Note that “simple CDS” is exactly the same as MCDS with numreads=1. There is no reason to use CDS.
Known modes are:
1 (or "Single") single sampling
2 (or "CDS") correlated double sampling
3 (or "MCDS") multiple-read correlated double sampling
4 (or "UTR") up the ramp samplingROTATOR
ROTATOR
FACSUM will always DISPLAY SCAM PA in position angle mode.
Observers want slit PA so set it in the GUI but remember that FACSUM shows SCAM PA
GINGA
The scam GINGA will have issues:
displays shifted images
does not overlay TAGS on new images
Creates strange patterns on the zoom window
In any case, it is best to set many exposures going with goi2 200
for example so that next images will balance out the issues.
NIRSPAO CONFIGURATION
set targwave
Typical offset and configuration
NIRSPAO + FIU
ERRORS
This is not good: