Table of Contents:


Relevant Devices

Description

Qty

Source

Part Number

Ion Pump Controller

4

Gamma

SPC-1-P-S-1-US110-E-S-N

Ion Pump

4

Gamma

3SDI1V5KNN (3S TiTan)

Air Cables

4

Gamma

10K-SC10-10K (10 m)

Vacuum Cables

4

Custom

Note there are two pumps per cryostat. One will be operated and the second will sit idle as a backup.

Each pump has a dedicated controller.


Relevant Functional Block Diagram

to be added


Software Actions

Action

Notes

1

Green controller 1: take pressure measurement

Value reported as telemetry (see below)

2

Green controller 2: take pressure measurement

Value reported as telemetry (see below)

3

Red controller 1: take pressure measurement

Value reported as telemetry (see below)

4

Red controller 2: take pressure measurement

Value reported as telemetry (see below)

5

Green controller 1: turn pump on

6

Green controller 1: turn pump off

7

Green controller 2: turn pump on

8

Green controller 2: turn pump off

9

Red controller 1: turn pump on

10

Red controller 1: turn pump off

11

Red controller 2: turn pump on

12

Red controller 2: turn pump off


Required Adjustable Settings

Setting

When Needed

Notes

Build

Observing

Maintenance


Required Telemetry

Parameter

Report Frequency

Purpose

Notes

1

Green controller 1 pressure value

Once per minute (TBC)

Monitor green cyrostat pressure

2

Green controller 2 pressure value

Once per minute (TBC)

Monitor green cyrostat pressure

3

Red controller 1 pressure value

Once per minute (TBC)

Monitor red cyrostat pressure

4

Red controller 2 pressure value

Once per minute (TBC)

Monitor red cyrostat pressure


Required Alarms

Parameter

Threshold

Notes

1

Green controller 1 pressure high

TBC

Only one pump per cyrostat will be in operation at any time.

The alarm system would need to know which pump/controller is in operation at any time to monitor the correct pressure value.

2

Green controller 2 pressure high

TBC

3

Red controller 1 pressure high

TBC

4

Red controller 2 pressure high

TBC


Relevant Datasheets & Manuals

Description

File

Source

1

SPC Controller Spec Sheet

https://www.gammavacuum.com/index.php/product?id=64

2

SPC Controller Manual

https://www.gammavacuum.com/index.php/product?id=64

3

SPC ModBus Register Map

https://www.gammavacuum.com/index.php/product?id=64

4

SPC Firmware

https://www.gammavacuum.com/index.php/product?id=64

5

SPC Labview Drivers
(likely not useful, but included for completeness' sake)

https://www.gammavacuum.com/index.php/product?id=64


Relevant Software

Description

File

Source

Snippit from device manual explaining simplicity of telnet connection.

SG tested this with PuTTY and it works; simplest command to test is the device name:

spc 01

which should return:

OK 00 SPC2

From page 14 of “SPC Controller Manual”, posted just above.

Full command list starts on page 19 of same manual.

(Steve is annotating the command list for KPF’s needs)

Python code that Steve lifted from web, slightly modified, and found it worked!

Key part in here (I think) is the command structure; this is the 'Read Pressure' command:

s.send(('spc 0B' + "\r\n").encode())

https://gist.github.com/tspspi/779d96f305f719082d2db1e201856375/revisions


Installation Instructions

Description

File

Enable the high pressure resolution mode (this is a front-panel-only setting).

  • No labels