Introduction

The optical bench within the vacuum chamber is supported by five vibration isolation pedestals. Typically these isolators have a hand-crank that can be used to adjust the height of each isolator mechanism and therefore level the table being supported.

For KPF, these five hand-cranks were replaced by vacuum-compatible stepper motors. This allows for remote leveling of the bench via computer control (while the bench is installed within the vacuum chamber).

...

Relevant Devices

Description

Qty

Source

Part Number

Stepper Motor Controller

1

Galil

DMC-4050(16BIT)-C012-I000-I000-D4040-D4020(ISAMP)

Controller Power Supply

1

Galil

PSR-12-24

Vacuum-Compatible Stepper Motors

5

Lin Engineering

4118M-06S-DDRO

...