Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  • Power off the ion pumps

  • modify -s kpfgreen COLDHEADRMO=on COLDHEADRMP=1.5

  • modify -s kpfgreen STA_CCDRMO=on STA_CCDRMP=1

    • The ramp units are degrees per minute

  • modify -s kpfgreen COLDHEADTRG=40 STA_CCDTRG=40

CCR Cool Down Procedure

  • modify -s kpfgreen COLDHEADRMO=on COLDHEADRMP=1.5

  • modify -s kpfgreen STA_CCDRMO=on STA_CCDRMP=1

  • modify -s kpfgreen COLDHEADTRG=-110 STA_CCDTRG=-100

  • When temperatures get to …?

  • modify -s kpfgreen COL_PUMPON=Yes