Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  • Set the ramp rates:

    • modify -s kpfgreen COLDHEADRMO=on COLDHEADRMP=1.5

    • modify -s kpfgreen STA_CCDRMO=on STA_CCDRMP=1

  • Start the cold head cooling:

    • modify -s kpfgreen COLDHEADTRG=-110

  • Wait a few minutes to let the CH start cooling, then:

    • modify -s kpfgreen STA_CCDTRG=-100

  • Once everything is down to temp or levels off because of the higher partial pressure in the cryostat (maybe then give it another 2 hours ?), and turn on the ion pump:

    • modify -s kpfgreen COL_PUMPON=Yes

...