to put the dichroic in as pwfs@vm-pwfsserver
prtcInsertDichroic
check with show -s ao obdb
and it should show ~220. mm
Modulator:
must be between 0 and 1V
...
so the position must be 0< | center + amplitude| < 1
scale is ~ 1V/arcsec
KPIC DIAGRAM
CRED2 Display
...
(Tracking Camera)
Bring up an xterm: pwfs@vm-pwfsserver (from the k2aoserver-new VNC pull down)
2. ssh -X nfiudev@nfiuserver
(the usual AO password for VNC)
3. [nfiudev@nfiuserver ~]$ cd src/sce
4 [nfiudev@nfiuserver sce]$ TrackCam tint 0 .002
[nfiudev@nfiuserver sce]$ ./psfDispTiny.py -s /tmp/TC_Display_images.im.shm
Another option is to bring up the VNC where the FIU camera is running an eavesdrop on that:
nfiuserver:2 <usual AO password>
...
PyWFS Quick Access
Launch with
...
open the loops
center modulation
PyWFS AO screens: k2aoserver-new:2
...
Instructions for Restarting the PyWFS camera (Saffira): (from Sylvain)
Symptom: PyWFS Camera (in the PyWFS status) shows red and says it is powered off
Clicking on it and rebooting/restarting the camera server does NOT resolve this.
Solution:
Connect to the RTC: ssh -X prtc@prtcserver
showRTProcess: it opens a window and the 2nd process from the top is partially crash
Ctrl-c to restart the process
Then connect to: ssh -X pwfs@vm-pwfsserver
Issue command: prtcRestartCOM (takes about 5 sec, then you should be able to reacquire the camera)