...
Table of Contents:
Table of Contents |
---|
Page Status
...
Date
...
Initials
...
Change
...
April 20, 2021
...
SG
...
Introduction
Two vacuum gauges are located on the main KPF vacuum chamber and are monitored by a Leybold Graphix Two gauge controller.
...
Relevant Devices
Description | Qty | Source | Part Number | Notes |
---|---|---|---|---|
Vacuum Gauge Controller | 1 | Leybold |
Graphix Two | Controls the gauges on the chamber | |||
High Vac gauge | 1 | Leybold | PTR 90 N | Combo Pirani and Cold Cathode gauge |
Low Vac gauge | 1 | Leybold | TTR 91 N S | Pirani gauge |
RS-232 to Ethernet Adapter | 1 | Lantronix | UDS 2100 | This is the version with two serial ports |
Relevant Functional Block Diagram
...
(This device at lower right)
Software Actions
...
Action
...
Notes
...
Read Channel 1 Pressure Value
...
Value reported as telemetry (see below)
...
Read Channel 2 Pressure Value
...
Value reported as telemetry (see below)
...
Channel 1 Gauge Status
...
Value reported as telemetry (see below)
...
Channel 2 Gauge Status
...
Value reported as telemetry (see below)
...
Set Point 1 Status
...
Value reported as telemetry (see below)
Required Adjustable Settings
...
Setting
...
When Needed
...
Notes
...
Build
...
Observing
...
Maintenance
...
Telemetry Read Cadence
...
Yes
...
No
...
Maybe
Device Set-up Parameters
View file | ||
---|---|---|
|
Required Telemetry
...
Parameter
...
Report Frequency
...
Purpose
...
Notes
...
Channel 1 Pressure Value
...
Once per minute [TBD]
...
Channel 2 Pressure Value
...
Once per minute [TBD]
...
Channel 1 Gauge Status
...
Once per minute [TBD]
...
Channel 2 Gauge Status
...
Once per minute [TBD]
...
Set Point 1 Status
...
Once per minute [TBD]
...
Check on gate valve position
Required Alarms
Parameter
Threshold
Notes
Chamber pressure too low
TBD
(but standard 9-pin RS-232 connectors) | ||||
Smaller Footprint Power Supply for Lantronix | 1 | Tri-Mag, LLC | Tri-Mag PN: L6R12H-120 DigiKey PN: 364-1264-ND | The stock Lantronix power supplies do not work well with the PDUs (as they block two extra outlets). KPF chose these replacement supplies since they only take up one PDU outlet. |
...
Functional Block Diagram
...
Cable Details
Cable | Qty | Details | Notes |
---|---|---|---|
Ethernet | 1 | Cat6 Copper Ethernet Cable M-M, 6 feet long | Entirely within Vac Cart. |
RS-232 | 1 | DB9, M-F, Straight through, 3 feet long | Entirely within Vac Cart. |
Gauge Cables (Vac Cart) | 2 | Shielded ethernet cables, RJ-45, M-M, 6 feet long | Entirely within Vac Cart. |
Gauge Cables (Chamber) | 2 | Shielded ethernet cables, RJ-45, M-M, 10 feet long | |
Bulkhead Coupler (Dual) | 1 | Universal Sub-Panel Black, 2 Black Feed-Thru Couplers, RJ45K (8x8K) Straight |
...
Pictures of Devices
...
Relevant Datasheets & Manuals
Description
File
Source
Description | File | Source | |||||
---|---|---|---|---|---|---|---|
1 | Leybold Graphix Two Manual |
| https://www.leyboldproducts.com/media/pdf/0f/9e/dc/300550402_002_C1_GRAPHIX_123_EN.pdf | ||||
2 |
Relevant Software
‘Vacuum Measurement’ brochure containing specs on various gauges, including both of our gauges (PTR 90 N and TTR 91 N S) |
| ||||||
3 | PTR 90 N Vacuum Gauge ‘Technical Data’ sheet |
| |||||
4 | CAD Model for PTR 90 N gauge. |
| |||||
5 | Datasheet for TTR 91 N S gauge. Seems a bit outdated, and was downloaded from a vac supply company (not Leybold). However there was nothing on this part number at Leybold (closest was a TTR RNS model, which might be the newer version of our gauge). |
| http://vacshop.ir/wp-content/uploads/2016/02/vacshop-THERMOVAC_New-TTR91N.pdf | ||||
6 | Lantronix UDS 2100 Quick Start Guide |
| https://www.lantronix.com/products/uds2100/#tab-docs-downloads | ||||
7 | Lantronix UDS 2100 User Guide |
| |||||
8 | Smaller Footprint Power Supply Datasheet |
| https://www.tri-mag.com/wp-content/uploads/2021/05/L6R12H_Series_2021-01.pdf (Linked off the DigiKey product page) |