Power off the ion pumps
modify -s kpfgreen COL_PUMPON=No
modify -s kpfgreen ECH_PUMPON=No
Set the lakeshore to ramping control with a rate:
Ensure ramping is on:
show -s kpfgreen STA_CCDRMO COLDHEADRMO
should both be “On”modify -s kpfgreen COLDHEADRMO=on COLDHEADRMP=1.5
modify -s kpfgreen STA_CCDRMO=on STA_CCDRMP=1
The ramp units are degrees per minute
Change the target temperature
modify -s kpfgreen COLDHEADTRG=40 STA_CCDTRG=40
...