Table of Contents:


Relevant Devices

Description

Qty

Source

Part Number

Lamp Intensity Monitor Controller

1

Newport

1830-R-GPIB

Lamp Intensity Monitor Detector

1

Newport

818-SL/DB

RS-232 to Ethernet Adapter

1

Lantronix

UDS2100

This is the version with two serial ports (and standard 9-pin RS-232 connectors)


Relevant Functional Block Diagram

Most recent version. Going forward, lamp intensity monitor and its connections are not expected to change.


Software Actions

Action

Notes

1

Single measurement.

That is, report the Wattage value the detector is currently reporting.

Note calibration unit translation stage has to move detector into beam first.

2

Calculate average of ‘n' discrete single measurements taken ’t' seconds apart.

Average several single measurements together to smooth out measurement noise/detector variability.

This value reported as telemetry (see below).

This approach was used for measurements at SSL using a similar device; see example Python code at bottom of this page.


Required Adjustable Settings

NOTE: device automatically switches from “LOCAL” to “REMOTE” mode when it receives a remote command through either the USB, RS-232C, or IEEE 488 GPIB interfaces.

Setting

When Needed

Notes

Build

Observing

Maintenance

Number of measurements 'n' to include in average.

YES

NO

MAYBE

Perhaps start with 10

Time delay 't' between measurements used for average.

YES

NO

MAYBE

Perhaps start with 0.5 seconds.

Device setup parameters follow:

Set offset subtraction

YES

NO

MAYBE

Equivalent to “ZERO” button on device.

KPF default should be ‘Z0’ (no offset subtraction).

Set calibration wavelength

YES

NO

MAYBE

Equivalent to wavelength buttons on device.

KPF default value should be 658 nm. Remote command might be 'W0658'.

Set attenuator mode

YES

NO

MAYBE

“ATTN” button on device.

KPF default should be set to ‘off’ (i.e. no attenuator in place). Remote command is ‘A0’

Set measurement units

YES

NO

MAYBE

Equivalent to “UNITS” button on device.

KPF default should be Watts (remote command ‘U1').

Set averaging mode

YES

NO

MAYBE

Equivalent to “AVG” button on device.

KPF default should be set to ‘Fast’ mode (no averaging by device). Remote command for ‘Fast’ mode is ‘F3’.

Set audible beep on/off

YES

NO

MAYBE

Equivalent to “BEEP” button on device.

KPF default should be ‘off’ (remote command ‘B0’)

Set automatic range

YES

NO

MAYBE

Equivalent to “AUTO” range on device.

KPF default should be auto-range (remote command ‘R0’).

Turn display on/off

YES

NO

MAYBE

KPF default should be ‘off’ (remote command ‘K0’). Likely device produces less heat (and isn’t a light source) with the display off.

Local lockout

YES

NO

MAYBE

I presume KPF default should be ‘locked’ while under remote operation (remote command ‘L1’)


Required Telemetry

Parameter

Report Frequency

Purpose

Notes

1

Source 1 intensity

TBC

Monitor calibration source outputs over time

Report this value if Source 1 is selected at octagon.

2

Source 2 intensity

TBC

Report this value if Source 2 is selected at octagon.

3

Source 3 intensity

TBC

Report this value if Source 3 is selected at octagon.

4

Source 4 intensity

TBC

Report this value if Source 4 is selected at octagon.

5

Source 5 intensity

TBC

Report this value if Source 5 is selected at octagon.

6

Source 6 intensity

TBC

Report this value if Source 6 is selected at octagon.

7

Source 7 intensity

TBC

Report this value if Source 7 is selected at octagon.

8

Source 8 intensity

TBC

Report this value if Source 8 is selected at octagon.


Required Alarms

Parameter

Threshold

Notes

1

Calibration Source 1 intensity low

TBD

2

Calibration Source 2 intensity low

TBD

3

Calibration Source 3 intensity low

TBD

4

Calibration Source 4 intensity low

TBD

5

Calibration Source 5 intensity low

TBD

6

Calibration Source 6 intensity low

TBD

7

Calibration Source 7 intensity low

TBD

8

Calibration Source 8 intensity low

TBD


Relevant Datasheets & Manuals

Description

File

Source

1

1830-R-GPIB Quick Start

https://www.newport.com/p/1830-R-GPIB

2

1830-R-GPIB Manual

https://www.newport.com/p/1830-R-GPIB

3

Controller Computer Interface Software

https://www.newport.com/p/1830-R-GPIB

4

Controller Firmware

https://www.newport.com/p/1830-R-GPIB

5

818-SL/DB Detector Manual

https://www.newport.com/p/818-SL--DB

6

Lantronix UDS 2100 Quick Start Guide

https://www.lantronix.com/products/uds2100/#tab-docs-downloads

7

Lantronix UDS 2100 User Guide

https://www.lantronix.com/products/uds2100/#tab-docs-downlo

Relevant Software

Description

Files

Notes

Example python code from VPH measurement system at SSL. Written for a Newport 1936-C intensity meter, which perhaps uses similar code to the KPF intensity meter (Newport 1830-R-GPIB)


Dispatcher Implementation

Keyword

Value

Type

CALMON_POW

Power measured by detector every time delay (see above)

Double - read only

CALMON_AVGPOW

Power measured by detector averaged over a number of steps specified in NAVG

Double - read only

CALMON_AVGTIME

Time stamp of last average measurement

Time in ascii format or unix timestamp in binary format - read only

CALMON_NAVG

Number of POW measurements to average to compute AVGPOW. Dispatcher requires > 1.

Integer - writable as as well readable, default set in configuration file

CALMON_AVG

Averaging is on or off.

Boolean

CALMON_CONN

Connection on or off to the device

Boolean

CALMON_IP

IP address of the Lantronix

String - read only but set in configuration file.

CALMON_PORT

Port id on the Lantronix for the serial connection

String - read only but set in configuration file.

CALMON_ID

ID string returned by device. Only valid when there is a connection.

String - read only.

CALMON_ATT

Attenuator on / off

Enum - 0 Off - 1 On

CALMON_BEEP

Device Beep on / off

Enum - 0 Off - 1 On

CALMON_DISPLAY

Front panel display on / off

Enum - 0 Off - 1 On

CALMON_FILTER

Averaging on the device

Enum - 1 Slow (16) - 2 Medium (4) - 3 Fast (1 or no averaging)

CALMON_LOCK

Front paneled locked remotely, on locks

Enum - 0 Off - 1 On

CALMON_RANGE

The device does automatic scaling of the output (0) or has eight preset ranges

Enum - 0 Auto - 1 2 3 4 5 6 7 8

CALMON_UNITS

The output units of the power measurement. Can be relative (REL) to a previous value.

Enum - 1 Watts - 2 dB - 3 dBm - 4 REL

CALMON_ZERO

Use a previous value as a zero point, can be turned on or off.

Enum - 0 Off - 1 On

CALMON_WAVE

Wavelength of light used for the detector calibration, specified in nanometers.

Integer - dispatcher enforces range between 0 and 2000

DISPSTA

Dispatcher status.

Enum and read only. If value is 4 or “Not Connected” the connection to the Vaisala or the Lantonix is not available. If value is 3 or “Connecting” the dispatcher is attempting to connect to the Vaisala through the Lantronix.

Configuration

Value

Explanation

ip

192.168.23.102

IP address of the Lantronix

port

10002

Port that is used, 10002 means port 2

poll_time

0.5

Time between data requests for the measured power value.

navg

10

Default starting value for NAVG. Keyword can be written.