Relevant Devices

Description

Qty

Source

Part Number

Lamp Intensity Monitor Controller

1

Newport

1830-R-GPIB

Lamp Intensity Monitor Detector

1

Newport

818-SL/DB

RS-232 to Ethernet Adapter

1

Lantronix

UDS2100

This is the version with two serial ports (and standard 9-pin RS-232 connectors)

Smaller Footprint Power Supply for Lantronix

1

Tri-Mag, LLC

Tri-Mag PN: L6R12H-120

DigiKey PN: 364-1264-ND

The stock Lantronix power supplies do not work well with the PDUs (as they block two extra outlets). KPF uses these replacement supplies since they only take up one PDU outlet.


Relevant Functional Block Diagram


Cable Details

Cable

Qty

Details

Notes

Ethernet

1

Cat6 Copper Ethernet Cable M-M, 6 feet long

Entirely within Cal Rack.

DB9

DB15 Male-Female Detector Extension Cable

BNC F-M Extension Cable

12 feet

McMaster-Carr

6641T35

50-Ohms, 22-Gauge


Pictures of Devices


Relevant Datasheets & Manuals

Description

File

Source

1

1830-R-GPIB Quick Start

https://www.newport.com/p/1830-R-GPIB

2

1830-R-GPIB Manual

https://www.newport.com/p/1830-R-GPIB

3

Controller Computer Interface Software

https://www.newport.com/p/1830-R-GPIB

4

Controller Firmware

https://www.newport.com/p/1830-R-GPIB

5

818-SL/DB Detector Manual

https://www.newport.com/p/818-SL--DB

6

818-SL-L Detector Calibration Report

Shipped with actual KPF device

7

Lantronix UDS 2100 Quick Start Guide

https://www.lantronix.com/products/uds2100/#tab-docs-downloads

8

Lantronix UDS 2100 User Guide

https://www.lantronix.com/products/uds2100/#tab-docs-downlo

9

Smaller Footprint Power Supply Datasheet

https://www.tri-mag.com/wp-content/uploads/2021/05/L6R12H_Series_2021-01.pdf

(Linked off the DigiKey product page)


Dispatcher Configuration File Details

Configuration

Value

Explanation

dipsnum

1

Dispatcher number

ip

192.168.23.178

IP address of the Lantronix

port

10001

Port that is used, 10001 means port 1

Lantronix serial config

19200, 8N1, No flow control

Serial port settings, matched to Newport 1830-R setting.

poll_time

0.5

Time between data requests for the measured power value.

navg

10

Default starting value for NAVG. Keyword can be written.

wave

658

Default starting value for the wave parameter, the value is nm.

units

1

Default starting value for units, this is Watts.

filter

3

Default starting value for the filter, this is the Fast value.

display

0

Default starting value for the display, this is the off value.


Calibration Intensity Measurement

Software Actions

The calibration intensity monitor measures the intensity of calibration lamps when it is in the beam. This dispatcher automates some of the configuration and writes the measured power to a keyword whose name matches the calibration source when the measurement being performed.

The measurement only occurs when the AVG value of the calibration intensity monitor is On. It also checks that the lamp is on for which ever source is selected by the octagon stage, and that the monitor is in the beam. If either state changes during a measurement, the keyword is no longer updated. Currently it does not check the calibration intensity monitor itself.

Dispatcher Implementation

Keyword

Value

Type

MEASURING

Yes / No - Is the average power being recorded.

boolean

INPUT

Octagon position selected, that is also the name of the keyword that the power will be recorded to.

enumerated

<LAMP_NAME>

The last average power value recorded.

Double

<LAMP_NAME>_TIME

The time that the last power value was recorded.

Double


Data taken with cal unit closed up June 7, 2022.

Wavelengths selected based on central band pass of the two detectors.

Dark was taken with the Solar calibrator unplugged and the intensity monitor out of the beam and the lamps were all off.

Source

Power at 522 nm (Watts)

Power at 735 nm (Watts)

Broad Band

1..72e-05

1.303e-05

UNe Daily

1.442e-05

1.063e-05

ThAr Daily

1.214e-05

9.011e-06

ThAr Gold

1.581e-05

1.153e-05

UNe Gold

2.044e-05

1.508e-05

Dark

7.116e-10

6.056e-10

Broad band on, detector at -15000

1.15e-9

Broad band on, detector at -25000

3.33e-9

Broad band on, detector at -35000

3.77e-8