Introduction

This is the CCD detector for the Calcium HK spectrometer.

The exposure time for this detector will be determined by the exposure meter system, and will not be known in advance. As such, this detector will be run in an ‘external exposure (bulb)’ mode that will allow an exposure to start when the trigger-in signal goes high, and stop the exposure when the trigger-in signal goes low. This mode is required as it allows the image to be read out after the trigger-in signal goes low.

This USB-connected device will be located within the AO Electronics room of Keck I. However, to decrease the heat load within AO on telescope, the computer it connects to will be located within the KI server room. To connect the two we use a USB-to-fiber optic link running from AO to the K1 server room. Note this scheme has the heritage of another Andor detector with same USB-to-fiber system already deployed at Keck and working fine (K2 ACAM).


Relevant Devices

Description

Qty

Source

Part Number

Ca HK Detector

1

Andor

Newton DU920P-BU-99C1
Final Quote:

USB Fiber Extender overview
(separate components listed below)

1

Adnacom

https://adnacom.com/product/s3b-order/

USB-to-Fiber Adapter

1

Adnacom

R1USB30B-03

Fiber

1

Adnacom

10-m purchased for lab tests. Final, on-summit cable already installed through az-wrap at Keck.

Fiber-to-PCIe Adapter

1

Adnacom

H1A-02-LP


Relevant Functional Block Diagram


Relevant Hardware Datasheets & Manuals

Description

File

Source

1

Andor Newton Manual

Email from Andor sales rep (Mike Zemlan)

2

Adnaco-S3B Overview Datasheet

https://adnacom.com/s3b/

3

Performance Benchmarking for Adnacom USB extensions

4

Adnaco-S3B Quick Start Guide

5

How to disable PCIe power management in Windows 7, 8 and 10

6

Adnaco-USB30B xHCI Filter Driver for Win8/8.1/10 (driver version 1.0.0.5) – sllc448a.zip

7

Adnaco-USB30B xHCI Driver for WinXP/Vista/7 (driver version 1.16.6.0) – sllc423k.zip

8

Adnaco-R1USB30 data sheet

9

Adnaco-H1A data sheet

10

Adnaco Gen2 Systems User’s Guide

11

Keck drawing for USB-to-fiber adapter in use for K2 ACAM

Ben McCarney (WMKO)


Relevant Software

Description

File

Source

1

Capability list for our Newton

(terse and verbose versions)

Output from Will Deich’s query program

2

Andor SDK 2 Manual

Note SDK 2 is what we need as it’s for CCD detectors

(SDK 3 is for CMOS detectors).

Direct link for these file sent by Andor sales rep (Mike Zemlan)