Overview: We want to keep power on to the red and blue ccd controllers in order to maintain stable temperatures during the power shutdown. Since the risk is overcooling from the loss of heaters, we will preemptively set the heaters to -115 C to get an extra 5 C buffer in case the power shutdown is extended.

Assumption: The whole instrument is on UPS power with only the red/blue detector controllers on in order to run the heaters.

...

Shut down power to everything other than the blue saddle bag and red archon:

  • Power down CCDs

    • From lrismonored > modify -s lredccd ccdpower=off

    • From manuka > modify -s lrisblue ccdpow=0

  • OPEN Instrument Tech > Start VNC Viewers

  • PULLDOWN Workspace Menu in VNC Viewer. SELECT: LRIS Engineering > LRIS Power Control.

  • On ‘LRIS Power Control’ GUI > Power off all components except blue saddlebag and red ccd archon

    Power down CCDs

  • From lrismonored > modify -s lredccd ccdpower=off

  • From manuka > modify -s lrisblue ccdpow=0

Proceed to shutdown the servers in the following order:

...