Scratchpad for procedures being developed/iterated on.
Close Gate Valve
Roots Pump | Turbo Pump | Gate Valve | Cart Pressure | Chamber Pressure | ||
---|---|---|---|---|---|---|
1 | Starting point Gate valve open, pumps running, and chamber under vacuum. | ON | ON | OPEN | 10-5 Torr | 10-5 Torr |
2 | Log into vacuum pump webserver On a computer on the KPF private network, open a web browser and go to 192.168.23.125. At top right of page, log in using the kpfdev account credentials. Remain on the ‘Main’ page. | ON | ON | OPEN | 10-5 Torr | 10-5 Torr |
3 | Find PDU and Outlet for gate valve power supply On the same computer, open another web browser tab/window and go to the PDU Outlet Registry. Find the gate valve on this page and note:
| ON | ON | OPEN | 10-5 Torr | 10-5 Torr |
4 | Log into PDU webserver On same computer, open another web browser tab/window and go to the IP address for the gate valve PDU. Log in using the kpfdev account credentials. Navigate to the ‘Outlet’ page shown in hierarchy at top left of screen. | ON | ON | OPEN | 10-5 Torr | 10-5 Torr |
5 | Close gate valve On the PDU webserver, turn off power to the appropriate outlet for the gate valve. This closes the gate valve. | ON | ON | CLOSED | Starts to fall towards 10-8 Torr | Starts to rise |
6 | Turn off turbo pump Now go back to the webserver page for the vacuum pump. Turn off the vacuum pump using the red ‘STOP' button at bottom center of page. Note pump will not instantaneously stop, but the pump frequency will slowly start to drop. The user has no control here, this process is internally regulated by the pump. Note it might take 1.5 hours for it to gracefully slow from 1000 Hz to 0 Hz. | ON | OFF | CLOSED | Once pump frequency slows down below 500 Hz the pressure will start to rise. | Rising |
7 | Ending point The turbo has stopped rotating and is showing a frequency of 0 Hz, and a power of 0 W. Roots pump is still running, and has reached a steady pressure of 1.5 x 10-2 Torr. | ON | OFF | CLOSED | 1.5 x 10-2 Torr | Rising |