LN2 Based Procedures from Kodi


CCR Warm Up Procedure

See day log K1-35087

  • Power off the ion pumps

  • modify -s kpfgreen COLDHEADRMO=on COLDHEADRMP=1.5

  • modify -s kpfgreen STA_CCDRMO=on STA_CCDRMP=1

    • The ramp units are degrees per minute

  • modify -s kpfgreen COLDHEADTRG=40 STA_CCDTRG=40

CCR Cool Down Procedure

  • modify -s kpfgreen COLDHEADRMO=on COLDHEADRMP=1.5

  • modify -s kpfgreen STA_CCDRMO=on STA_CCDRMP=1

  • modify -s kpfgreen COLDHEADTRG=-110

  • Wait a few minutes to let the CH start cooling

  • modify -s kpfgreen STA_CCDTRG=-100

  • When temperatures get to …?

  • modify -s kpfgreen COL_PUMPON=Yes