No additional power cycling capabilities needed.

Power Outage Procedure

  • Shut down instrument software in preparation for powering things off

    • Run CSUSTOP

      • modify -s mcsus stopcsu=1

    • Ensure detector is put to bed gently

      • modify -s mds abort=1

      • modify -s mds powerdown=1

      • mosfire stop mds

  • Go to low power mode by turning off components in the electronics cabinet

    • Power down these components (e.g. using the Power Control GUI):

      • CSU Controller

      • CSU Drives

      • Jade 2

      • Computer

      • Ne lamp

      • Ar lamp

      • Guider focus galil (these prevent frost on guider, so probably need to leave these on?)

      • Guider camera electronics (these prevent frost on guider, so probably need to leave these on?)

      • Motor box

      • FCS controller

    • This leaves telemetry active (needs mosfireserver and vm-mosfire)

      • Could shut down Kaimana if desperate