Vacuum Chamber Vent Valve

Vacuum Chamber Vent Valve


Table of Contents:


Introduction

A vent valve is located on the door flange of the main vacuum chamber. It has a solenoid within that is controlled by mains power: the vent valve opens when power is applied, and closes when power is removed.

Within KPF it is remotely-controlled by turning its PDU outlet on or off (via kpfpower).

This valve should only be operated by engineering staff while following the chamber venting procedure.


Relevant Devices

Description

Qty

Source

Part Number

Description

Qty

Source

Part Number

Vent Valve for NW 25 with 115V Solenoid

1

Agilent Varian

L9942302


Relevant Functional Block Diagram


Cable Details

Cable

Qty

Details

Cable

Qty

Details

Vent Solenoid Cable (120V)

1

NEMA 5-15P cable spliced onto pigtail leads from vent valve.
Cable length is likely 6 feet (but might be 10 feet).


Relevant Datasheets & Manuals

Description

File

Source

Description

File

Source

1

Relevant Section of Agilent Varian Valve Catalog (with our valve highlighted).

https://www.idealvac.com/files/literature/11_Varian_2011_Vacuum_Valves.pdf

2

Full Agilent Varian Valve Catalog